Clinical validation of the non-invasive cardiac output monitor USCOM-1A in critically ill patients
Authors:
- A. R. H. van ZantenAZA. R. H. van Zanten
- G. F. BormGBG. F. Borm
- D. H. T. TjanDTD. H. T. Tjan
- 1 November 2008
- journal article
- Published by Wolters Kluwer Health in European Journal of Anaesthesiology
- Vol. 25 (11) , 917-924
Abstract
D and gaining attention. To evaluate the accuracy of the ultrasonic cardiac output monitor (USCOM)-1A (Pty Ltd, Coffs Harbour, NSW, Australia), a non-invasive cardiac output device including its capability to differentiate between different shock states in haemodynamically unstable ICU patients was used in this single-centre, prospective, observational study. Methods: Cardiac output was measured with a pulmonary artery catheter and transcutaneously via a suprasternal approach with the USCOM-1A by continuous-wave Doppler ultrasound in 25 adult patients in a mixed medical and surgical ICU in a major teaching hospital in the Netherlands. Results: A total of 1315 USCOM-1A cardiac output measurements were performed. In order to reduce time-variability, the mean of five consecutive USCOM-1A measurements was calculated. Total 263 values were compared with 263 thermodilution cardiac output measurements performed with a pulmonary artery catheter. Data were analysed for systematic error, precision and correlation. Systematic and random errors were found. On average USCOM-1A values were 12% lower than thermodilution measurements (systematic error), while the random error was 17% (coefficient of variation). The error comprised an inter-operator variability of 3%, an inter-patient variability of 11% and residual variability of 15%. The correlation coefficient of the calculated cardiac index with the USCOM-1A and the pulmonary artery catheter was r = 0.8024 and 0.6438, respectively. Temperature and gender did not influence correlations. The learning curve for USCOM-1A skill acquisition was steep. Conclusions: The correlation between the two techniques was acceptable, although relevant systematic and variable errors were detected. USCOM-1A provided adequate data to distinguish non-invasively different shock types in ICU patients....Keywords
This publication has 26 references indexed in Scilit:
- Non-Invasive Cardiac Output Determination by Two-Dimensional Independent Doppler During and After Cardiac SurgeryThe Annals of Thoracic Surgery, 2005
- Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomised controlled trialThe Lancet, 2005
- Influence analysis for linear mixed‐effects modelsStatistics in Medicine, 2004
- Early Use of the Pulmonary Artery Catheter and Outcomes in Patients With Shock and Acute Respiratory Distress SyndromeA Randomized Controlled TrialJAMA, 2003
- A randomised, controlled trial of the pulmonary artery catheter in critically ill patientsIntensive Care Medicine, 2002
- The Pulmonary Artery Catheter—Friend, Foe, or Accomplice?JAMA, 2001
- The Effectiveness of Right Heart Catheterization in the Initial Care of Critically III PatientsJAMA, 1996
- New perspectives in the assessment of cardiac chamber dimensions during development and adulthoodJACC, 1992
- Interobserver agreement and accuracy of bedside estimation of right and left ventricular ejection fraction in acute myocardial infarctionThe American Journal of Cardiology, 1989
- CLINICAL CALORIMETRYArchives of internal medicine (1908), 1916