Abstract
A simple and accurate ultra-high-performance liquid chromatography tandem mass spectrometry (UHPLC-MS/MS) method was developed and validated for the first time as a confirmatory method for the simultaneous determination of stilbenes – hexestrol and diethylstibestrol in serum. Extraction was based on a simple acid denaturation of protein followed by liquid–liquid extraction using methyl tert butyl ether. Extracts were directly injected into the UHPLC-MS/MS without further purification. Excellent recoveries in the range 82–99% and 91–128% were obtained for hexestrol and diethylstibestrol, respectively. Both within-day repeatability and between-day reproducibility were generally satisfactory with RSD 0.99 for both hexestrol and diethylstibestrol. Both matrix effect and uncertainties associated with sample preparation and instrumental analysis were significantly reduced with the use of a deuterated compound (hexestrol-d4) as internal standard. The LOD and LOQ were 0.09 and 0.08 ng/ml, and 0.28 and 0.25 ng/ml, respectively, for hexestrol and diethylstibestrol. The method was found to be suitable for the simultaneous determination of hexestrol and diethylstibestrol in serum.
Funding Information
  • the Agricultural Research Council under Residue Project

This publication has 32 references indexed in Scilit: