Improved 3D correction for partial volume effects in brain SPECT
Authors:
- 1 January 1997
- journal article
- research article
- Published by Wiley in Human Brain Mapping
- Vol. 5 (5) , 379-388
Abstract
An improved method for correction of partial volume effects (PVE) in brain SPECT is proposed. It is fully three-dimensional, does not require particular patient positioning, and works with scans only partially covering the brain. The location of functionally inactive brain regions (primarily cerebrospinal fluid) is extracted from high-resolution MRI. An automatic 3D registration algorithm then determines the geometric transformation between MRI and SPECT. Correction consists of: 1) counting the volumetric active/inactive ratio in each volume element of the functional scan using the measured SPECT point spread function; 2) correcting the functional measures according to these ratios; 3) fusing functional and anatomical information at the resolution of MRI. Quantitative validation was performed using a phantom containing a test region in which multiple parallel acrylic plates thinner than SPECT resolution created high PVE, as well as a large reference region not suffering from PVE. Reference activity was recovered in the test region with an accuracy of 1–3%. The method was applied to clinical images demonstrating a combination of hypoperfusion and cortical atrophy. The composite anatomical-functional corrected images, in which the main sulci are visible, yield better differentiation between decreased function and focal atrophy. Hum. Brain Mapping 5:379–388, 1997.Keywords
This publication has 14 references indexed in Scilit:
- Robust multimodality registration for brain mappingHuman Brain Mapping, 1997
- Positron Emission Tomography Metabolic Data Corrected for Cortical Atrophy Using Magnetic Resonance ImagingAlzheimer Disease & Associated Disorders, 1996
- Brain single‐photon emission computed tomography *Neurology, 1994
- 3-D SPECT simulations of a complex 3-D mathematical brain model: effects of 3-D geometric detector response, attenuation, scatter, and statistical noiseIEEE Transactions on Medical Imaging, 1992
- Quantitative analysis in single photon emission tomography (SPET)European Journal of Nuclear Medicine and Molecular Imaging, 1992
- Correction of PET Data for Partial Volume Effects in Human Cerebral Cortex by MR ImagingJournal of Computer Assisted Tomography, 1990
- Correction of Positron Emission Tomography Data for Cerebral AtrophyJournal of Cerebral Blood Flow & Metabolism, 1986
- Quantitation in Positron Emission Computed Tomography: 5. Physical–Anatomical EffectsJournal of Computer Assisted Tomography, 1981
- Quantitation in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1979
- Two Methods for Calculating Regional Cerebral Blood Flow from Emission Computed Tomography of Inert Gas ConcentrationsJournal of Computer Assisted Tomography, 1979