Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy
Authors:
Open Access
- 14 December 2021
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 23 (2) , e13501
Abstract
Purpose: A novel, mobile cone‐beam computed tomography (CBCT) system for image‐guided adaptive brachytherapy was recently deployed at our hospital as worldwide first site. Prior to the device's clinical operation, a profound characterization of its imaging performance was conducted. This was essential to optimize both the imaging workflow and image quality for achieving the best possible clinical outcomes. We present the results of our investigations.Methods: The novel CBCT‐system features a ring gantry with 121 cm clearance as well as a 43.2 × 43.2 cm2 flat‐panel detector, and is controlled via a tablet‐personal computer (PC). For evaluating its imaging performance, the geometric reproducibility as well as imaging fidelity, computed tomography (CT)‐number accuracy, uniformity, contrast‐noise‐ratio (CNR), noise characteristics, and spatial resolution as fundamental image quality parameters were assessed. As dose metric the weighted cone‐beam dose index (CBDIw) was measured. Image quality was evaluated using standard quality assurance (QA) as well as anthropomorphic upper torso and breast phantoms. Both in‐house and manufacturer protocols for abdomen, pelvis, and breast imaging were examined.Results: Using the in‐house protocols, the QA phantom scans showed altogether a high image quality, with high CT‐number accuracy (R2 > 0.97) and uniformity (<12 Hounsfield Unit (HU) cupping), reasonable noise and imaging fidelity, and good CNR at bone–tissue transitions of up to 28:1. Spatial resolution was strongly limited by geometric instabilities of the device. The breast phantom scans fulfilled clinical requirements, whereas the abdomen and pelvis scans showed severe artifacts, particularly at air/bone–tissue transitions.Conclusion: With the novel CBCT‐system, achieving a high image quality appears possible in principle. However, adaptations of the standard protocols, performance enhancements in image reconstruction referring to artifact reductions, as well as the extinction of geometric instabilities are imperative.Keywords
This publication has 35 references indexed in Scilit:
- A quality assurance framework for the fully automated and objective evaluation of image quality in cone‐beam computed tomographyMedical Physics, 2014
- Combining scatter reduction and correction to improve image quality in cone‐beam computed tomography (CBCT)Medical Physics, 2010
- Geometric accuracy of digital volume tomography and conventional computed tomographyBritish Journal of Oral and Maxillofacial Surgery, 2008
- What is Cone-Beam CT and How Does it Work?Dental Clinics of North America, 2008
- C-arm Cone-beam CT: General Principles and Technical Considerations for Use in Interventional RadiologyJournal of Vascular and Interventional Radiology, 2008
- Imaging doses from the Elekta Synergy X-ray cone beam CT systemBritish Journal of Radiology, 2007
- Intraoperative cone‐beam CT for guidance of head and neck surgery: Assessment of dose and image quality using a C‐arm prototypeMedical Physics, 2006
- Accurate technique for complete geometric calibration of cone‐beam computed tomography systemsMedical Physics, 2005
- Implementation of a cone‐beam reconstruction algorithm for the single‐circle source orbit with embedded misalignment correction using homogeneous coordinatesMedical Physics, 2001
- Practical cone-beam algorithmJournal of the Optical Society of America A, 1984