Optimal caliper placement: manual vs automated methods
Open Access
- 13 May 2013
- journal article
- research article
- Published by Wiley in Ultrasound in Obstetrics & Gynecology
- Vol. 43 (2) , 170-175
Abstract
Objective: To examine the inter‐ and intraoperator repeatability of manual placement of calipers in the assessment of basic fetal biometric measurements and to compare the results with those of an automated caliper placement system.Methods: We used stored ultrasound images of 95 normal fetuses between 19 and 25 weeks' gestation. Five operators (two experts, one resident and two students) were asked to measure the biparietal diameter (BPD), occiptofrontal diameter (OFD), abdominal circumference (AC) and femur length (FL) twice manually and twice automatically. For each operator, the repeatability of manual and automated measurements was assessed using intraoperator SD. For the assessment of interoperator repeatability, the mean of the four manual measurements by the two experts for each fetus was used as the gold standard. The relative bias of the manual measurements of the three non‐expert operators and the operator‐independent automated measurement were compared with the gold standard measurement by mean and 95% CI.Results: In 89.5% of the 95 cases, the automated measurement algorithm was able to obtain appropriate measurements of BPD, OFD, AC and FL. Intraoperator SDs for the manual measurements ranged between 0.15 and 1.56, irrespective of the experience of the operator. For the automated biometric measurement system, there was no difference between the measurements of each operator. Regarding interoperator repeatability, the mean difference between the manual measurements of the two students, the resident and the gold standard was between −0.10 and 2.53 mm. The automated measurements tended to be closer to the gold standard, but the difference in bias in automated vs manual measurements did not reach statistical significance.Conclusion: In about 90% of cases, it was possible to obtain basic biometric measurements with an automated system. The use of automated measurements resulted in a significant improvement in intraoperator repeatability, but measurements were not significantly closer to the gold standard of expert examiners. Copyright © 2013 ISUOG. Published by John Wiley & Sons Ltd.Keywords
This publication has 17 references indexed in Scilit:
- What are the limits of accuracy in fetal weight estimation with conventional biometry in two‐dimensional ultrasound? A novel postpartum studyUltrasound in Obstetrics & Gynecology, 2012
- Intra‐ and interobserver variability in fetal ultrasound measurementsUltrasound in Obstetrics & Gynecology, 2011
- Accuracy of sonographic fetal weight estimation of fetuses with a birth weight of 1500 g or lessEuropean Journal of Obstetrics & Gynecology and Reproductive Biology, 2010
- Performance of 36 Different Weight Estimation Formulae in Fetuses with MacrosomiaFetal Diagnosis and Therapy, 2010
- A New Sonographic Weight Formula for Fetuses ≤ 2500 gUltraschall in der Medizin - European Journal of Ultrasound, 2009
- Gender-Specific Weight Estimation of Fetuses between 2,501 and 3,999 g – New Regression FormulaeFetal Diagnosis and Therapy, 2008
- An integrated model for the prediction of preeclampsia using maternal factors and uterine artery Doppler velocimetry in unselected low-risk womenAmerican Journal of Obstetrics and Gynecology, 2005
- Intraobserver and interobserver reproducibility of fetal biometryUltrasound in Obstetrics & Gynecology, 2004
- A systematic review of the ultrasound estimation of fetal weightUltrasound in Obstetrics & Gynecology, 2004
- Sonographic estimation of fetal weight. The value of femur length in addition to head and abdomen measurements.Radiology, 1984