A rapid and sensitive 384‐well microtitre format chemiluminescent enzyme immunoassay for 19‐nortestosterone
- 6 January 2003
- journal article
- research article
- Published by Wiley in Luminescence
- Vol. 18 (2) , 72-78
Abstract
We developed a competitive chemiluminescent (CL) enzyme immunoassay for rapid, sensitive analysis of 19‐nortestosterone (19‐NT) in bovine urine. Anti‐19‐NT polyclonal antibodies were raised in rabbits using a 19‐NT‐hemisuccinate derivative conjugated with ovalbumin; the derivative was also conjugated with horseradish peroxidase (HRP) as a label. Antibodies were immobilized on 384‐well black polystyrene microtitre plates and HRP‐labelled 19‐NT activity was measured using an efficient chemiluminescent substrate (SuperSignal® ELISA Femto) after 3 min incubation. Emitted light was recorded using a conventional, photomultiplier‐tube‐based microtitre plate reader or a sensitive back‐illuminated, cooled CCD camera. The developed method fulfils all the requirements of precision (intra‐ and inter‐assay CV < 10%) and accuracy (mean recovery 94–112%), with a detection limit of 0.03 ppb (1.1 × 10−9 mol/L) in a urine matrix. Chemiluminescence enhances detectability of the HRP‐labelled tracer (thus lowering the limit of detection with respect to colorimetry) and reduces analysis time. The 384‐well microtitre plate cuts the sample/reagent volume (20 µL), a five‐fold reduction with respect to the conventional 96‐well microtitre plate. The developed method is suitable for high‐throughput screening of 19‐NT in urine samples, with reduced costs as compared with conventional colorimetric enzyme immunoassays. Copyright © 2003 John Wiley & Sons, Ltd.Keywords
This publication has 17 references indexed in Scilit:
- Chemiluminescent Imaging of Enzyme-Labeled Probes Using an Optical Microscope–Videocamera LuminographAnalytical Biochemistry, 1998
- Residues and metabolism of 19-nortestosterone laurate in steers†The Analyst, 1998
- Chemiluminescent Low-Light Imaging of Biospecific Reactions on Macro- and Microsamples Using a Videocamera-Based LuminographAnalytical Chemistry, 1996
- Screening of calf urine for 19-nortestosterone: matrix effect in some immunoassaysThe Analyst, 1995
- Enzyme immunoassay screening procedure for the synthetic anabolic estrogens and androgens diethylstilbestrol, nortestosterone, methyltestosterone and trenbolone in bovine urineJournal of Chromatography B: Biomedical Sciences and Applications, 1989
- Effective monitoring of residues of nortestosterone and its major metabolite in bovine urine and bileJournal of Chromatography B: Biomedical Sciences and Applications, 1989
- A Specific Radioimmunoassay for the Detection of 19-Nortestosterone Residues in Urine and Plasma of CattleJournal of Immunoassay, 1986
- Development of a chemiluminescent immunoassay for 19‐nortestosterone (nandrolone)Veterinary Quarterly, 1984
- Production of a high-titer antibody to bile acidsThe Journal of Steroid Biochemistry and Molecular Biology, 1980
- Chemical Coupling of Peptides and Proteins to Polysaccharides by Means of Cyanogen HalidesNature, 1967