Relative Sensitivity Factors for Submicron Secondary Ion Mass Spectrometry with Gallium Primary Ion Beam
Authors:
- Hitomi SatohHSHitomi Satoh
- Masanori OwariMOMasanori Owari
- Yoshimasa NiheiYNYoshimasa Nihei
- 1 August 1993
- journal article
- Published by IOP Publishing in Japanese Journal of Applied Physics
- Vol. 32 (8R)
Abstract
Relative sensitivity factors (RSFs) of thirteen elements in the oxide glass matrix in secondary ion mass spectrometry (SIMS) excited by a gallium focused ion beam were determined. RSFs were obtained by analyzing powder particles of standard glass samples. Whole volumes of each particles were analyzed in the “shave-off” mode in order to avoid topographic effects. Reproducibility of RSFs was good, and sample-to-sample scattering of values was relatively small. Dependence of RSFs on the first ionization potential was shown to be reasonable. In order to check the validity of the RSFs, coal fly ash particles were analyzed. The results were in reasonable agreement with the data obtained through the bulk chemical analysis.Keywords
This publication has 4 references indexed in Scilit:
- Relative sensitivity factors and useful yields for a microfocused gallium ion beam and time‐of‐flight secondary ion mass spectrometerJournal of Vacuum Science & Technology A, 1991
- Quantitative analysis by submicron secondary ion mass spectrometryJournal of Vacuum Science & Technology B, 1988
- High spatial resolution secondary ion mass spectrometry with parallel detection systemJournal of Vacuum Science & Technology A, 1987
- Scanning ion microscopy: Elemental maps at high lateral resolutionApplied Surface Science, 1986