B1 mapping for bias‐correction in quantitative T1 imaging of the brain at 3T using standard pulse sequences
- 16 March 2017
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 46 (6) , 1673-1682
Abstract
Purpose: B1 mapping is important for many quantitative imaging protocols, particularly those that include whole‐brain T1 mapping using the variable flip angle (VFA) technique. However, B1 mapping sequences are not typically available on many magnetic resonance imaging (MRI) scanners. The aim of this work was to demonstrate that B1 mapping implemented using standard scanner product pulse sequences can produce B1 (and VFA T1) maps comparable in quality and acquisition time to advanced techniques.Materials and Methods: Six healthy subjects were scanned at 3.0T. An interleaved multislice spin‐echo echo planar imaging double‐angle (EPI‐DA) B1 mapping protocol, using a standard product pulse sequence, was compared to two alternative methods (actual flip angle imaging, AFI, and Bloch‐Siegert shift, BS). Single‐slice spin‐echo DA B1 maps were used as a reference for comparison (Ref. DA). VFA flip angles were scaled using each B1 map prior to fitting T1; the nominal flip angle case was also compared.Results: The pooled‐subject voxelwise correlation (ρ) for B1 maps (BS/AFI/EPI‐DA) relative to the reference B1 scan (Ref. DA) were ρ = 0.92/0.95/0.98. VFA T1 correlations using these maps were ρ = 0.86/0.88/0.96, much better than without B1 correction (ρ = 0.53). The relative error for each B1 map (BS/AFI/EPI‐DA/Nominal) had 95th percentiles of 5/4/3/13%.Conclusion: Our findings show that B1 mapping implemented using product pulse sequences can provide excellent quality B1 (and VFA T1) maps, comparable to other custom techniques. This fast whole‐brain measurement (∼2 min) can serve as an excellent alternative for researchers without access to advanced B1 pulse sequences.Level of Evidence: 1Technical Efficacy: Stage 1J. Magn. Reson. Imaging 2017;46:1673–1682.Keywords
Funding Information
- Natural Sciences and Engineering Research Council of Canada
- CIHR
This publication has 39 references indexed in Scilit:
- Transmit field inhomogeneity and T1estimation errors in breast DCE‐MRI at 3 teslaJournal of Magnetic Resonance Imaging, 2013
- Nonlinear formulation of the magnetic field to source relationship for robust quantitative susceptibility mappingMagnetic Resonance in Medicine, 2012
- Robust and Fast Whole Brain Mapping of the RF Transmit Field B1 at 7TPLOS ONE, 2012
- Mapping of low flip angles in magnetic resonancePhysics in Medicine & Biology, 2011
- Optimization and validation of methods for mapping of the radiofrequency transmit field at 3TMagnetic Resonance in Medicine, 2010
- B1 mapping by Bloch‐Siegert shiftMagnetic Resonance in Medicine, 2010
- MP2RAGE, a self bias-field corrected sequence for improved segmentation and T1-mapping at high fieldNeuroImage, 2010
- Rapid radiofrequency field mapping in vivo using single‐shot STEAM MRIMagnetic Resonance in Medicine, 2008
- Quantitative magnetization transfer imaging in postmortem multiple sclerosis brainJournal of Magnetic Resonance Imaging, 2007
- A new look at the method of variable nutation angle for the measurement of spin-lattice relaxation times using fourier transform NMRJournal of Magnetic Resonance (1969), 1977