Noninvasive Control of the Power Transferred to an Implanted Device by an Ultrasonic Transcutaneous Energy Transfer Link
Authors:
- 1 April 2014
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 61 (4) , 995-1004
Abstract
Ultrasonic transcutaneous energy transfer is an effective method for powering implanted devices noninvasively. Nevertheless, the amount of power harvested by the implanted receiver is sensitive to the distance and orientation of the external transmitting transducer attached to the skin with respect to the implanted receiving transducer. This paper describes an ultrasonic power transfer link whose harvested power is controlled by an inductive link. A small (5 μF) storage capacitor voltage, which is part of the implanted unit, is allowed to swing between 3.8 and 3.5 V using hysteretic control. The two control states are indicated by excitation (while the implanted storage capacitor voltage decreases) or the absence of excitation of an implanted coil that is magnetically coupled to an external coil attached to the skin surface. A 35 mW Ultrasonic Transcutaneous Energy Transfer link was fabricated using two piezoelectric transducers of equal size (Fuji Ceramics C-2 PZT disc 15 mm × 3 mm) operated at a vibration frequency of 720 kHz. By applying the proposed hysteretic control, the captured power was effectively regulated for implantation depths of up to 85 mm.Keywords
Funding Information
- Israel Ministry of science (3-7962)
- Taiwan Scientific Cooperation Program
This publication has 16 references indexed in Scilit:
- Ultra-low-power wireless implantable blood flow sensing microsystem for vascular graft applications2011
- Contactless energy transfer through air by means of ultrasound2011
- Wireless Implantable Electronic Platform for Chronic Fluorescent-Based BiosensorsIEEE Transactions on Biomedical Engineering, 2011
- Energy Efficiency and Reliability in Wireless Biomedical Implant SystemsIEEE Transactions on Information Technology in Biomedicine, 2011
- Ultrasonic transcutaneous energy transfer using a continuous wave 650kHz Gaussian shaded transmitterUltrasonics, 2010
- Self-Sampled All-MOS ASK Demodulator for Lower ISM Band ApplicationsIEEE Transactions on Circuits and Systems II: Express Briefs, 2010
- Modeling and Optimization of Printed Spiral Coils in Air, Saline, and Muscle Tissue EnvironmentsIEEE Transactions on Biomedical Circuits and Systems, 2009
- Transcutaneous powering of implantable micro-stimulators for functional restoration of impaired gastrointestinal motility2004
- Frequency-dependent skin-effect formulation for resistance and internal inductance of a solid cylindrical conductorIEE Proceedings - Microwaves, Antennas and Propagation, 2004
- New equivalent circuits for elementary piezoelectric transducersElectronics Letters, 1970