Radiofrequency coils for magnetic resonance microscopy
Authors:
- 26 February 2008
- journal article
- review article
- Published by Wiley in NMR in Biomedicine
- Vol. 22 (9) , 975-981
Abstract
Given the several orders of magnitude fewer spins per voxel for MR microscopy than for conventional MRI, efficient coil design is important to obtain sufficient signal‐to‐noise within reasonable data acquisition times. As MR microscopy is typically performed using very high magnetic fields, coil design must also incorporate the effects of increased component losses and skin‐depth‐dependent resistance, as well as radiation losses and phase effects for coils when conductor dimensions constitute a substantial fraction of the electromagnetic wavelength. For samples much less than 1 mm in size, wire solenoids or microfabricated planar coils are used. For samples with diameters of several millimeters, saddle, birdcage, Alderman–Grant or millipede coils become the preferred choice. Recent advances in multiple‐coil probes and phased arrays have been used to reduce data acquisition time and/or increase sample throughput, and small superconducting coils have shown significant improvements in signal‐to‐noise over equivalently sized room‐temperature coils. Copyright © 2008 John Wiley & Sons, Ltd.Keywords
This publication has 58 references indexed in Scilit:
- Diffusion-limited resolution in nuclear magnetic resonance microscopyJournal of Magnetic Resonance (1969), 2004
- Optimizing the point spread function in phase‐encoded magnetic resonance microscopyConcepts in Magnetic Resonance Part A, 2004
- MR microscopy and high resolution small animal MRI: applications in neuroscience researchProgress in Neurobiology, 2002
- Solenoidal microcoil design—Part II: Optimizing winding parameters for maximum signal‐to‐noise performanceConcepts in Magnetic Resonance, 2001
- Solenoidal microcoil design. Part I: Optimizing RF homogeneity and coil dimensionsConcepts in Magnetic Resonance, 2001
- Radiofrequency microcoils in magnetic resonanceProgress in Nuclear Magnetic Resonance Spectroscopy, 1997
- Correlated Susceptibility and Diffusion Effects in NMR Microscopy Using both Phase-Frequency Encoding and Phase-Phase EncodingJournal of Magnetic Resonance, Series B, 1994
- Susceptibility-limited resolution in nuclear magnetic resonance microscopyJournal of Magnetic Resonance (1969), 1990
- Sensitivity and resolution in NMR imagingJournal of Magnetic Resonance (1969), 1987
- The signal-to-noise ratio of the nuclear magnetic resonance experimentJournal of Magnetic Resonance (1969), 1976