“TBDx” is an innovative smear microscopy system that automatically loads slides onto a microscope, focuses and digitally captures images and then classifies smears as positive or negative using computerised algorithms. To determine the diagnostic accuracy of TBDx, using culture as the gold standard, and compare this to a microscopist's diagnostic performance. This study is nested within a cross-sectional study of tuberculosis suspects from South African gold mines. All tuberculosis suspects had one sputum sample collected, which was decontaminated prior to smear microscopy, liquid culture and organism identification. All slides were auramine-stained and then read by both a research microscopist and by TBDx using fluorescence microscopes, classifying slides based on the WHO classification standard of 100 fields of view (FoV) at 400× magnification. Of 981 specimens, 269 were culture positive for Mycobacterium tuberculosis (27.4%). TBDx had higher sensitivity than the microscopist (75.8% versus 52.8%, respectively), but markedly lower specificity (43.5% versus 98.6%, respectively). TBDx classified 520/981 smears (53.0%) as scanty positive. Hence, a proposed hybrid software/human approach that combined TBDx examination of all smears with microscopist re-examination of TBDx scanty smears was explored by replacing the “positive” result of slides with 1–9 AFB detected on TBDx with the microscopist's original reading. Compared to using the microscopist's original results for all 981 slides, this hybrid approach resulted in equivalent specificity, a slight reduction in sensitivity from 52.8% to 49.4% (difference of 3.3%; 95% confidence interval: 0.2%, 6.5%), and a reduction in the number of slides to be read by the microscopist by 47.0%. Compared to a research microscopist, the hybrid software/human approach had similar specificity and positive predictive value, but sensitivity requires further improvement. Automated microscopy has the potential to substantially reduce the number of slides read by microscopists.Keywords
This publication has 13 references indexed in Scilit:
- Genotype MTBDRplus for Direct Detection of Mycobacterium tuberculosis and Drug Resistance in Strains from Gold Miners in South AfricaJournal of Clinical Microbiology, 2012
- Thibela TB: Design and methods of a cluster randomised trial of the effect of community-wide isoniazid preventive therapy on tuberculosis amongst gold miners in South AfricaContemporary Clinical Trials, 2011
- Rapid Molecular Detection of Tuberculosis and Rifampin ResistanceNew England Journal of Medicine, 2010
- Liquid vs. solid culture for tuberculosis: performance and cost in a resource-constrained setting.2010
- Automated detection of tuberculosis in Ziehl‐Neelsen‐stained sputum smears using two one‐class classifiersJournal of Microscopy, 2009
- Comparison of four culture systems for Mycobacterium tuberculosis in the Zambian National Reference Laboratory.2009
- Image processing techniques for identifying Mycobacterium tuberculosis in Ziehl-Neelsen stains.2008
- Fluorescence versus conventional sputum smear microscopy for tuberculosis: a systematic reviewThe Lancet Infectious Diseases, 2006
- Optimal tuberculosis case detection by direct sputum smear microscopy: how much better is more?2002
- The importance of quality control of sputum smear microscopy: the effect of reading errors on treatment decisions and outcomes.1999