Comparison of methods for determining pulse arrival time from Doppler and photoplethysmography signals
- 1 August 2014
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 18 (1094687X) , 3809-3812
Abstract
The aim of this study was to compare three foot-finding methods applied to ultrasound Doppler and photoplethysmographic (PPG) signals: maximum 1st derivative, maximum 2nd derivative and an `intersecting tangents' method. The pulse arrival times of each method were compared. Also the precision of each method was evaluated by comparing instantaneous cardiac periods derived using each method from simultaneous Doppler and PPG with a reference measurement: the R-R interval calculated from a simultaneously recorded ECG. The results show that the maximum 1st derivative method produced significantly larger pulse arrival times than the other two methods. The intersecting tangents method produced greatest precision for cardiac periods compared with ECG than maximum 1st or 2nd derivatives for both Doppler (r2 = 0.975) and PPG (r2 = 0.987) signals.Keywords
This publication has 11 references indexed in Scilit:
- Photoplethysmography pulse rate variability as a surrogate measurement of heart rate variability during non-stationary conditionsPhysiological Measurement, 2010
- Determinants of pulse wave velocity in healthy people and in the presence of cardiovascular risk factors: ‘establishing normal and reference values’European Heart Journal, 2010
- Comparative study of methodologies for pulse wave velocity estimationJournal of Human Hypertension, 2008
- Vascular StiffnessClinical Journal of the American Society of Nephrology, 2008
- Assessment of heart rate variability derived from finger-tip photoplethysmography as compared to electrocardiographyJournal of Medical Engineering & Technology, 2008
- Can Photoplethysmography Variability Serve as an Alternative Approach to Obtain Heart Rate Variability Information?Journal of Clinical Monitoring and Computing, 2007
- Association of Risk Factors With Increased Pulse Wave Velocity Detected by a Novel Method Using Dual-Channel PhotoplethysmographyAmerican Journal of Hypertension, 2005
- Validation of a device to measure arterial pulse wave velocity by a photoplethysmographic methodPhysiological Measurement, 2002
- Determination of pulse wave velocities with computerized algorithmsAmerican Heart Journal, 1991
- Non-invasive estimate of the mechanical properties of peripheral arteries from ultrasonic and photoplethysmographic measurementsClinical Physics and Physiological Measurement, 1991