Porosity of the wall of a Neurolac® nerve conduit hampers nerve regeneration
Authors:
- 1 January 2009
- journal article
- research article
- Published by Wiley in Microsurgery
- Vol. 29 (6) , 473-478
Abstract
One way to improve nerve regeneration and bridge longer nerve gaps may be the use of semipermeable/porous conduits. With porosity less biomaterial is used for the nerve conduit. We evaluated the short‐term effects of porous Neurolac® nerve conduits for in vivo peripheral nerve regeneration. In 10 male Black Hooded rats, a gap of 10 mm was bridged by a porous Neurolac® nerve conduit. Evaluation point ranged from 3 to 12 weeks. The sciatic nerve function was not measurable due to automutilation and flexion contractures. The gait‐stance duration showed no improvement with time, indicating a disturbed walking pattern. The nerve guides showed very fast degradation with swelling, fragmentation, and collapse. Furthermore, a severe foreign body reaction occurred. Nerve regeneration was severely hampered. This study showed no beneficial effects of porous Neurolac® nerve conduits when compared with previous findings with nonporous copolymeric nerve guides of a slightly different composition. © 2009 Wiley‐Liss, Inc. Microsurgery, 2009.Keywords
This publication has 18 references indexed in Scilit:
- US Food and Drug Administration/Conformit Europe-Approved Absorbable Nerve Conduits for Clinical Repair of Peripheral and Cranial NervesAnnals of Plastic Surgery, 2008
- Effects of unidirectional permeability in asymmetric poly(DL‐lactic acid‐co‐glycolic acid) conduits on peripheral nerve regeneration: An in vitro and in vivo studyJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2007
- Clinical Use of Nerve Conduits in Peripheral-Nerve Repair: Review of the LiteratureJournal of Reconstructive Microsurgery, 2002
- The Role of Conduits in Nerve Repair: A ReviewReviews in the Neurosciences, 1996
- Gait-stance duration as a measure of injury and recovery in the rat sciatic nerve modelJournal of Neuroscience Methods, 1994
- High molecular weight copolymers of l-lactide and ?-caprolactone as biodegradable elastomeric implant materialsPolymer Bulletin, 1991
- Rat Sciatic Nerve Regeneration within an Acrylic Semipermeable Tube and Comparison with a Silicone Impermeable MaterialJournal of Neuropathology & Experimental Neurology, 1990
- Functional Evaluation of Complete Sciatic, Peroneal, and Posterior Tibial Nerve Lesions in the RatPlastic and Reconstructive Surgery, 1989
- Blind-ended semipermeable guidance channels support peripheral nerve regeneration in the absence of a distal nerve stumpBrain Research, 1988
- Permeable tubes increase the length of the gap that regenerating axons can spanBrain Research, 1987