Effects of Ultrasound Section-Thickness on Brachytherapy Needle Tip Localization Error
- 1 January 2011
- book chapter
- Published by Springer Nature
- Vol. 14, 299-306
Abstract
Purpose: Ultrasound section-thickness is the out-of-plane beamwidth causing major roles in creating image artifacts normally appearing around the anechoic areas. These artifacts can introduce errors in localizing the needle tips during any ultrasound-guided procedure. To study how section-thickness and imaging parameters can affect observing and localizing needle tips, we have conducted a typical calibration setup experiment. Method: Multiple needles were inserted orthogonal to the axial image plane, at various distances from the transducer. The experiment was conducted on a brachytherapy stepper for a curvilinear transrectal-ultrasound probe. Result: Experiments demonstrated that the imaging parameters have direct impacts on observing needle tips at different axial locations. They suggest specific settings to minimize the imaging artifacts. Conclusion: The ultrasound section-thickness and side lobes could result in misjudgment of needle insertion depth in an ultrasound-guided procedure. A beam profile could assist in considering the likelihood of position errors, when the effects of side lobes are minimized.Keywords
This publication has 6 references indexed in Scilit:
- Section-thickness profiling for brachytherapy ultrasound guidance2011
- A real-time ultrasound calibration system with automatic accuracy control and incorporation of ultrasound section thickness2008
- Test Object for Measurement of Section Thickness at USRadiology, 1999
- Estimation of ultrasound beam width in the elevation (section thickness) plane.Radiology, 1991
- The importance of ultrasonic side-lobe artifacts.Radiology, 1982
- Slice‐thickness artifacts in gray‐scale ultrasoundJournal of Clinical Ultrasound, 1981