Computational Design and Experimental Performance Testing of an Axial-Flow Pediatric Ventricular Assist Device

Authors:
  • DLD Scott Lim
    D Scott Lim
  • MMMichael A. McCulloch
    Michael A. McCulloch
  • WJWei Jiang
    Wei Jiang
  • XSXinwei Song
    Xinwei Song
  • PAPaul E. Allaire
    Paul E. Allaire
  • HWHouston G. Wood
    Houston G. Wood
  • DODon B. Olsen
    Don B. Olsen

Abstract
The Virginia Artificial Heart Institute continues to design and develop an axial-flow pediatric ventricular assist device (PVAD) for infants and children in the United States. Our research team has created a database to track potential PVAD candidates at the University of Virginia Children's Hospital. The findings of this database aided with need assessment and design optimization of the PVAD. A numerical analysis of the optimized PVAD1 design (PVAD2 model) was also completed using computational fluid dynamics (CFD) to predict pressure-flow performance, fluid force estimations, and blood damage levels in the flow domain. Based on the PVAD2 model and after alterations to accommodate manufacturing, a plastic prototype for experimental flow testing was constructed via rapid prototyping techniques or stereolithography. CFD predictions demonstrated a pressure rise range of 36–118 mm Hg and axial fluid forces of 0.8–1.7 N for flows of 0.5–3 l/min over 7,000–9,000 rpm. Blood damage indices per CFD ranged from 0.24% to 0.35% for 200 massless and inert particles analyzed. Approximately 187 (93.5%) of the particles took less than 0.14 seconds to travel completely through the PVAD. The mean residence time was 0.105 seconds with a maximum time of 0.224 seconds. Additionally, in a water/glycerin blood analog solution, the plastic prototype produced pressure rises of 20–160 mm Hg for rotational speeds of 5,960 ± 18 rpm to 9,975 ± 31 rpm over flows from 0.5 to 4.5 l/min. The numerical results for the PVAD2 and the prototype hydraulic testing indicate an acceptable design for the pump, represent a significant step in the development phase of this device, and encourage manufacturing of a magnetically levitated prototype for animal experiments.

This publication has 13 references indexed in Scilit: