Computational Design and Experimental Performance Testing of an Axial-Flow Pediatric Ventricular Assist Device
Authors:
- D Scott LimDLD Scott Lim
- Michael A. McCullochMMMichael A. McCulloch
- Wei JiangWJWei Jiang
- Xinwei SongXSXinwei Song
- Paul E. AllairePAPaul E. Allaire
- Houston G. WoodHWHouston G. Wood
- Don B. OlsenDODon B. Olsen
- 1 September 2005
- journal article
- research article
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 51 (5) , 629-635
Abstract
The Virginia Artificial Heart Institute continues to design and develop an axial-flow pediatric ventricular assist device (PVAD) for infants and children in the United States. Our research team has created a database to track potential PVAD candidates at the University of Virginia Children's Hospital. The findings of this database aided with need assessment and design optimization of the PVAD. A numerical analysis of the optimized PVAD1 design (PVAD2 model) was also completed using computational fluid dynamics (CFD) to predict pressure-flow performance, fluid force estimations, and blood damage levels in the flow domain. Based on the PVAD2 model and after alterations to accommodate manufacturing, a plastic prototype for experimental flow testing was constructed via rapid prototyping techniques or stereolithography. CFD predictions demonstrated a pressure rise range of 36–118 mm Hg and axial fluid forces of 0.8–1.7 N for flows of 0.5–3 l/min over 7,000–9,000 rpm. Blood damage indices per CFD ranged from 0.24% to 0.35% for 200 massless and inert particles analyzed. Approximately 187 (93.5%) of the particles took less than 0.14 seconds to travel completely through the PVAD. The mean residence time was 0.105 seconds with a maximum time of 0.224 seconds. Additionally, in a water/glycerin blood analog solution, the plastic prototype produced pressure rises of 20–160 mm Hg for rotational speeds of 5,960 ± 18 rpm to 9,975 ± 31 rpm over flows from 0.5 to 4.5 l/min. The numerical results for the PVAD2 and the prototype hydraulic testing indicate an acceptable design for the pump, represent a significant step in the development phase of this device, and encourage manufacturing of a magnetically levitated prototype for animal experiments.Keywords
This publication has 13 references indexed in Scilit:
- Computational Analysis of an Axial Flow Pediatric Ventricular Assist DeviceArtificial Organs, 2004
- Quantitative Evaluation of Blood Damage in a Centrifugal VAD by Computational Fluid DynamicsJournal of Fluids Engineering, 2004
- Design of a Continuous Flow Centrifugal Pediatric Ventricular Assist DeviceThe International Journal of Artificial Organs, 2003
- Pediatric Circulatory Support SystemsASAIO Journal, 2002
- Computational Fluid Dynamics and Experimental Validation of a Microaxial Blood PumpASAIO Journal, 2001
- Test Controller Design, Implementation, and Performance for a Magnetic Suspension Continuous Flow Ventricular Assist DeviceArtificial Organs, 1999
- Characterization of a Magnetic Bearing System and Fluid Properties for a Continuous Flow Ventricular Assist DeviceArtificial Organs, 1999
- Performance of a Continuous Flow Ventricular Assist Device: Magnetic Bearing Design, Construction, and TestingArtificial Organs, 1998
- Three‐Dimensional Numerical Prediction of Stress Loading of Blood Particles in a Centrifugal PumpArtificial Organs, 1995
- Model for a General Mechanical Blood Damage PredictionArtificial Organs, 1995