Development of a high-speed and a high-sensitive laser scanning magneto-optical imaging system
Open Access
- 1 February 2009
- conference paper
- Published by IOP Publishing in Journal of Physics: Conference Series
- Vol. 150 (1) , 012029
Abstract
A high-speed and a high-sensitive laser scanning magneto-optical (MO) imaging system have been developed. In the high-speed imaging mode, we have succeeded in almost the real time observation of MO images with the sensitivity better than 100 μT. On the other hand, in the high-sensitive mode using an acoustic-optic modulator and a RF lock-in amplifier, it was found that the system noise reduced into the magnetic field strength was less than 6 μT at the data acquisition time of 10 μs. In this mode we have also succeeded in the fast MO imaging of the magnetic flux penetration into a square-shaped YBa2Cu3O7-δ n film at ~0.3 s/frame for 128x128 and ~5 s/frame for 500x500 pixels' image.Keywords
This publication has 6 references indexed in Scilit:
- DC and AC Responses of Josephson Vortex Flow Transistors with High Tc Superconducting Thin FilmsIEICE Transactions on Electronics, 2006
- Macroscopic Quantum Tunneling in a -Wave High- SuperconductorPhysical Review Letters, 2005
- Magneto-optical studies of current distributions in high-TcsuperconductorsReports on Progress in Physics, 2002
- Control of magnetic flux in a YBa2Cu3O7−δ thin film loop using femtosecond laser pulsesApplied Physics Letters, 1997
- Direct observation of the current distribution in thin superconducting strips using magneto-optic imagingPhysical Review B, 1996
- RSFQ logic/memory family: a new Josephson-junction technology for sub-terahertz-clock-frequency digital systemsIEEE Transactions on Applied Superconductivity, 1991