Multisite trial of MR flow measurement: Phantom and protocol design
Authors:
- Paul E. SummersPSPaul E. Summers
- Hristo N. NikolovHNHristo N. Nikolov
- Maria DrangovaMDMaria Drangova
Open Access
- 15 April 2005
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 21 (5) , 620-631
Abstract
Purpose To describe a portable, easily assembled phantom with well-defined bore geometry together with a series of tests that will form the basis of a standardized quality assurance protocol in a multicenter trial of flow measurement by the MR phase mapping technique. Materials and Methods The phantom consists of silicone polymer layers containing parallel straight and stenosed flow channels in one layer and a U-bend in a second layer, separated by hermetically sealed agarose slabs. The phantom is constructed by casting low melting-point metal in an aluminum mold precisely milled to the desired geometry, and then using the low melting-point metal core as a negative around which the silicone is allowed to set. By melting out the metal, the flow channels are established. The milled aluminum mold is reusable, ensuring faithful reproduction of the flow geometry for all phantoms thus produced. The agarose layers provide additional loading and static background signal for background correction. With the use of the described phantom, one can evaluate flow measurement accuracy and repeatability, as well as the influence of several imaging geometry factors: slice offset, in-plane position, and slice-flow obliquity. Results The new phantom is compact and portable, and is well suited for reassembly. We were able to demonstrate its facility in a battery of tests of interest in evaluating MR flow measurements. Conclusion The phantom is a robust standardized test object for use in a multicenter trial. Such a trial, to investigate the performance of MR flow measurement using the phantom and the tests we describe, has been initiated. J. Magn. Reson. Imaging 2005;21:620–631.Keywords
This publication has 54 references indexed in Scilit:
- Resolution-insensitive velocity and flow rate measurement in low-background phase-contrast MRAMagnetic Resonance in Medicine, 2004
- Combined analysis of spatial and velocity displacement artifacts in phase contrast measurements of complex flowsJournal of Magnetic Resonance Imaging, 1997
- 13th Annual Meeting European Society for Magnetic Resonance in Medicine and Biology Prague, September 12–15, 1996Magnetic Resonance Materials in Physics, Biology and Medicine, 1996
- Artifacts and signal loss due to flow in the presence of Bo inhomogeneityMagnetic Resonance in Medicine, 1996
- Role of imaging parameters in phase mapping flow measurementMagnetic Resonance Materials in Physics, Biology and Medicine, 1994
- Cine magnetic resonance imaging and color Doppler flow mapping displays of flow velocity, spatial acceleration, and jet formation: A comparative in vitro studyAmerican Heart Journal, 1993
- Quantification of aortic regurgitation by magnetic resonance velocity mappingAmerican Heart Journal, 1993
- Computer-controlled flow simulator for MR flow studiesJournal of Magnetic Resonance Imaging, 1992
- Encoding strategies for three‐direction phase‐contrast MR imaging of flowJournal of Magnetic Resonance Imaging, 1991
- The measurement of time-averaged flow by magnetic resonance imaging using continuous acquisition in the carotid arteries and its comparison with Doppler ultrasoundClinical Physics and Physiological Measurement, 1990