An optimized approach for increasing lesion size in temperature‐controled setting using a catheter with a surface thermocouple and efficient irrigation
Open Access
- 22 April 2024
- journal article
- research article
- Published by Wiley in Journal of Arrhythmia
- Vol. 40 (3) , 536-551
Abstract
Background: We explore an optimized approach for increasing lesion size using a novel ablation catheter with a surface thermocouple and efficient irrigation in a temperature‐control setting.Methods: We conducted radiofrequency applications at various power levels (35 W, 40 W, and 45 W), contact forces (CFs, 10 g/20 g), and durations (60 s/120 s/180 s) in perpendicular/parallel catheter orientations, with normal saline irrigation (NS‐irrigation) and Half NS‐irrigation (HNS‐irrigation) in an ex‐vivo model (Step 1). In addition, we performed applications (35 W/40 W/45 W for 60 s/120 s/180 s in NS‐irrigation and 35 W/40 W for 60 s/120 s/180 s in HNS‐irrigation) in four swine (Step 2), evaluating lesion characteristics and the occurrence of steam pops.Results: In Step 1, out of 288 lesions, we observed 47 (16.3%) steam pops, with 13 in NS‐irrigation and 34 in HNS‐irrigation (p = .001). Although steam pops were mostly observed with the most aggressive setting (45 W/180 s, 54%) with NS‐irrigation, they happened in less aggressive settings with HNS irrigation. Lesion size significantly increased with longer‐duration ablation but not with HNS‐irrigation. The optimal %impedance‐drop cutoff to predict steam pops was 20% with a negative‐predictive‐value (NPV) = 95.1% including NS‐ and HNS‐irrigation groups, and 22% with an NPV = 96.1% in NS‐irrigation group. In Step 2, similar to the ex‐vivo model, lesion size significantly increased with longer‐duration ablation but not with HNS‐irrigation. Steam pops were absent with NS‐irrigation (0/35) even with the largest %impedance‐drop reaching 31% at 45 W/180 s. All steam pops were observed with HNS‐irrigation (6/21, 29%). The optimal %impedance‐drop cutoff predicting steam pops was 24% with an NPV = 96.3% including both NS‐ and HNS‐irrigation groups.Conclusions: Rather than using HNS‐irrigation, very long‐duration of radiofrequency applications up to 45 W/180 s may be recommended to safely and effectively increase lesion dimensions using this catheter with NS‐irrigation.Keywords
Funding Information
- Japan Society for the Promotion of Science (20K17074, 22K16068)
This publication has 28 references indexed in Scilit:
- Safe and effective delivery of high-power, short-duration radiofrequency ablation lesions with a flexible-tip ablation catheterHeart Rhythm O2, 2022
- Impact of a formula combining local impedance and conventional parameters on lesion size predictionJournal of Interventional Cardiac Electrophysiology, 2021
- Evaluation of Radiofrequency Ablation Irrigation TypeJACC: Clinical Electrophysiology, 2020
- Temperature-Controlled Radiofrequency Ablation Using Irrigated CathetersJACC: Clinical Electrophysiology, 2020
- Prospective Multicenter Experience With Cooled Radiofrequency Ablation Using High Impedance Irrigant to Target Deep Myocardial Substrate Refractory to Standard AblationJACC: Clinical Electrophysiology, 2018
- Role of Contact Force Sensing in Catheter Ablation of Cardiac ArrhythmiasJACC: Clinical Electrophysiology, 2018
- Use of Tissue Electric and Ultrasound Characteristics to Predict and Prevent Steam-Generated Cavitation During High-Power Radiofrequency AblationJACC: Clinical Electrophysiology, 2018
- Radiofrequency Ablation Using an Open Irrigated Electrode Cooled With Half-Normal SalineJACC: Clinical Electrophysiology, 2017
- Clinical and biophysical evaluation of variable bipolar configurations during radiofrequency ablation for treatment of ventricular arrhythmiasHeart Rhythm, 2016
- Multicenter Outcomes for Catheter Ablation of Idiopathic Premature Ventricular ComplexesJACC: Clinical Electrophysiology, 2015