Interchangeability between Pentacam and IOLMaster in Phakic Intraocular Lens Calculation
- 10 November 2014
- journal article
- research article
- Published by SAGE Publications in European Journal of Ophthalmology
- Vol. 25 (3) , 202-207
Abstract
Purpose: To evaluate the agreement and interchangeability between IOLMaster and Pentacam in phakic intraocular lens (IOL) calculations. Methods: This was a prospective study involving measurement of the anterior segment parameters anterior chamber depth (ACD), white to white (WTW) line, and average anterior corneal curvature (K-reading) on 108 healthy eyes of 56 consecutive patients. Interdevice agreement and interchangeability in phakic IOL calculations was evaluated. Results: Pentacam measured significantly longer ACD measurements (3.54 ± 0.35 mm) than the IOLMaster (3.40 ± 0.37 mm) (p<0.0001); on the other hand, it measured significantly shorter WTW line measurements (11.61 ± 0.44 mm) than the IOLMaster (11.66 ± 0.42 mm) (p = 0.016). No difference was found regarding the average K-readings. For biometry and phakic IOL power calculation, there was no clinically significant difference between the devices, but for phakic IOL diameter calculation the difference was clinically significant. Conclusions: There is a good agreement in ACD and K-readings between Pentacam and IOLMaster, making them interchangeable in biometry and phakic IOL power calculation, but they are not interchangeable regarding WTW line measurement; the difference will affect the phakic IOL diameter, which would affect lens safety.Keywords
This publication has 26 references indexed in Scilit:
- Comparison of corneal curvature and anterior chamber depth measurements using the manual keratometer, lenstar LS 900 and the pentacamMiddle East African Journal of Ophthalmology, 2013
- Assessment of Anterior Chamber Depth Using Visante Optical Coherence Tomography, Slitlamp Optical Coherence Tomography, IOL Master, Pentacam and Orbscan IIzOphthalmologica, 2010
- Anterior Segment Parameters Using Pentacam and Prediction of Corneal Endothelial Cell Loss after Cataract SurgeryKorean Journal of Ophthalmology, 2010
- Comparison of horizontal corneal diameter measurements using Galilei, EyeSys and Orbscan II systemsClinical and Experimental Optometry, 2009
- Comparison of anterior chamber depth measurements taken with the Pentacam, Orbscan IIz and IOLMaster in myopic and emmetropic eyesActa Ophthalmologica, 2009
- Anterior Chamber Depth Measurement in Phakic and Pseudophakic EyesOptometry and Vision Science, 2008
- Comparison of Different Techniques of Anterior Chamber Depth and Keratometric MeasurementsAmerican Journal of Ophthalmology, 2007
- Comparison of manual and automated methods to determine horizontal corneal diameterJournal of Cataract & Refractive Surgery, 2004
- Anterior chamber depth measured by two methods in myopic and hyperopic phakic IOL implantBritish Journal of Ophthalmology, 2000
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986