Influence of balloon design, plaque material composition, and balloon sizing on acute post angioplasty outcomes: An implicit finite element analysis
- 17 May 2021
- journal article
- research article
- Published by Wiley in International Journal for Numerical Methods in Biomedical Engineering
- Vol. 37 (8) , e03499
Abstract
In this work we propose a generic modeling approach for simulating percutaneous transluminal angioplasty (PTA) endovascular treatment, and evaluating the influence of balloon design, plaque composition, and balloon sizing on acute post‐procedural outcomes right after PTA, without stent implantation. Clinically‐used PTA balloons were classified into two categories according to their compliance characteristics, and were modeled correspondingly. Self‐defined elastoplastic constitutive laws were implemented within the plaque and artery models, after calibration based on experimental and clinical data. Finite element method (FEM) implicit solver was used to simulate balloon inflation and deflation. Besides balloon profile at max inflation, results are mainly assessed in terms of the elastic recoil ratio (ERR) and lumen gain ratio (LGR) obtained immediately after PTA. No variations in ERR nor LGR values were detected when the balloon design changed, despite the differences observed in their profile at max inflation. Moreover, LGR and ERR inversely varied with the augmentation of calcification level within the plaque (−11% vs. +4% respectively, from fully lipidic to fully calcified plaque). Furthermore, results showed a direct correlation between balloon sizing and LGR and ERR, with noticeably higher rates of change for LGR (+18% and +2% for LGR and ERR respectively for a calcified plaque and a balloon pressure increasing from 10 to 14 atm). However a larger LGR comes with a higher risk of arterial rupture. This proposed methodology opens the way for evaluation of angioplasty balloon selections towards clinical procedure optimization.Keywords
Funding Information
- Agence Nationale de la Recherche (ANR‐11‐LABX‐0004)
- Région Bretagne
This publication has 49 references indexed in Scilit:
- Patient‐specific finite element analysis of carotid artery stenting: a focus on vessel modelingInternational Journal for Numerical Methods in Biomedical Engineering, 2012
- Site specific inelasticity of arterial tissueJournal of Biomechanics, 2012
- Inelasticity of Human Carotid Atherosclerotic PlaqueAnnals of Biomedical Engineering, 2011
- A Methodology to Analyze Changes in Lipid Core and Calcification Onto Fibrous Cap Vulnerability: The Human Atherosclerotic Carotid Bifurcation as an Illustratory ExampleJournal of Biomechanical Engineering, 2009
- A Numerical Model to Study the Interaction of Vascular Stents with Human Atherosclerotic LesionsAnnals of Biomedical Engineering, 2007
- Modeling Plaque Fissuring and Dissection during Balloon Angioplasty InterventionAnnals of Biomedical Engineering, 2007
- Multiaxial Mechanical Characteristics of Carotid PlaqueStroke, 2007
- Finite element analysis of the implantation of a balloon-expandable stent in a stenosed arteryInternational Journal of Cardiology, 2005
- Hyperelastic modelling of arterial layers with distributed collagen fibre orientationsJournal of The Royal Society Interface, 2005
- Comparison of acute elastic recoil after directional coronary atherectomy versus standard balloon angioplastyAmerican Heart Journal, 1992