Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density
Open Access
- 8 September 2016
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 17 (5) , 466-481
Abstract
Radiochromic film has become an important tool to verify dose distributions for intensity‐modulated radiotherapy (IMRT) and quality assurance (QA) procedures. A new radiochromic film model, EBT3, has recently become available, whose composition and thickness of the sensitive layer are the same as those of previous EBT2 films. However, a matte polyester layer was added to EBT3 to prevent the formation of Newton's rings. Furthermore, the symmetrical design of EBT3 allows the user to eliminate side‐orientation dependence. This film and the flatbed scanner, Epson Perfection V750, form a dosimetry system whose intrinsic characteristics were studied in this work. In addition, uncertainties associated with these intrinsic characteristics and the total uncertainty of the dosimetry system were determined. The analysis of the response of the radiochromic film (net optical density) and the fitting of the experimental data to a potential function yielded an uncertainty of 2.6%, 4.3%, and 4.1% for the red, green, and blue channels, respectively. In this work, the dosimetry system presents an uncertainty in resolving the dose of 1.8% for doses greater than 0.8 Gy and less than 6 Gy for red channel. The films irradiated between 0 and 120 Gy show differences in the response when scanned in portrait or landscape mode; less uncertainty was found when using the portrait mode. The response of the film depended on the position on the bed of the scanner, contributing an uncertainty of 2% for the red, 3% for the green, and 4.5% for the blue when placing the film around the center of the bed of scanner. Furthermore, the uniformity and reproducibility radiochromic film and reproducibility of the response of the scanner contribute less than 1% to the overall uncertainty in dose. Finally, the total dose uncertainty was 3.2%, 4.9%, and 5.2% for red, green, and blue channels, respectively. The above uncertainty values were obtained by minimizing the contribution to the total dose uncertainty of the film orientation and film homogeneity.PACS number(s): 87.53.BnKeywords
This publication has 43 references indexed in Scilit:
- Evaluation of an Epson flatbed scanner to read Gafchromic EBT films for radiation dosimetryPhysics in Medicine & Biology, 2009
- Analysis of uncertainties in Gafchromic® EBT film dosimetry of photon beamsPhysics in Medicine & Biology, 2008
- Radiation transmission, leakage and beam penumbra measurements of a micro‐multileaf collimator using GafChromic EBT filmJournal of Applied Clinical Medical Physics, 2008
- Radiochromic film dosimetry with flatbed scanners: A fast and accurate method for dose calibration and uniformity correction with single film exposureMedical Physics, 2008
- Multi-dimensional dosimetric verification of stereotactic radiotherapy for uveal melanoma using radiochromic EBT filmZeitschrift für Medizinische Physik, 2008
- Megavoltage x-ray skin dose variation with an angle using grid carbon fibre couch topsPhysics in Medicine & Biology, 2007
- Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurancePhysics in Medicine & Biology, 2007
- Verification of cell irradiation dose deposition using a radiochromic filmPhysics in Medicine & Biology, 2007
- Precautions and strategies in using a commercial flatbed scanner for radiochromic film dosimetryPhysics in Medicine & Biology, 2006
- Absorption spectra variations of EBT radiochromic film from radiation exposurePhysics in Medicine & Biology, 2005