Investigation of intracranial aneurysm hemodynamics following flow diverter stent treatment
Authors:
- Y. ZhangYZY. Zhang
- W. ChongWCW. Chong
- Y. QianYQY. Qian
- 9 August 2012
- journal article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 35 (5) , 608-615
Abstract
Flow diverters (FDs) are high density meshed stents designed to reduce blood flow into intra-cranial aneurysms. Though the FD is one of many intracranial aneurysm (IA) treatments, FD implantation may also result in the growth and rupture of residual aneurysms. The purpose of this research is to investigate the effect of FD implantation on IA hemodynamics. Computational fluid dynamics (CFD) was conducted to analyze dynamic and resistance forces after FD deployment. Simulation results for the successful case (patient A) showed that FD flow resistance force was higher than dynamic force. This indicated that the FD provided sufficient resistance to reduce flow into the aneurysm. As a result, flow velocity magnitude at the aneurysm neck was reduced by 95%. On the other hand, the flow velocity magnitude at the aneurysm neck was reduced by about 50% for the unsuccessful case (patient B). The reason was that the flow resistance force at the aneurysm neck section was calculated to be lower than the flow driving force. In order to completely occlude the aneurysm, a higher resistance FD stent is to be required to suppress the dynamic forces. Patient-specific hemodynamic simulations offer means of quantitative estimation FD treatment outcomes.Keywords
This publication has 18 references indexed in Scilit:
- Flow diverters for treatment of intracranial aneurysms: Current status and ongoing clinical trialsJournal of Clinical Neuroscience, 2011
- Hydrogel-coated coils versus bare platinum coils for the endovascular treatment of intracranial aneurysms (HELPS): a randomised controlled trialThe Lancet, 2011
- Early Experience in the Treatment of Intra-Cranial Aneurysms by Endovascular Flow Diversion: A Multicentre Prospective StudyPLOS ONE, 2010
- Numerical simulation of hemodynamics in stented internal carotid aneurysm based on patient-specific modelJournal of Biomechanics, 2010
- Clinical application of image-based CFD for cerebral aneurysmsInternational Journal for Numerical Methods in Biomedical Engineering, 2010
- CURATIVE ENDOVASCULAR RECONSTRUCTION OF CEREBRAL ANEURYSMS WITH THE PIPELINE EMBOLIZATION DEVICENeurosurgery, 2009
- The Asymmetric Vascular StentStroke, 2009
- Effects of stent porosity on hemodynamics in a sidewall aneurysm modelJournal of Biomechanics, 2008
- Computational fluid dynamics of stented intracranial aneurysms using adaptive embedded unstructured gridsInternational Journal for Numerical Methods in Fluids, 2007
- Unruptured intracranial aneurysms: natural history, clinical outcome, and risks of surgical and endovascular treatmentThe Lancet, 2003