Transit Time Difference Flowmeter for Intravenous Flow Rate Measurement Using 1–3 Piezoelectric Composite Transducers
- 14 July 2017
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Sensors Journal
- Vol. 17 (17) , 5741-5748
Abstract
The flow rate of injected medication implemented by intravenous (IV) systems must be accurately monitored and meticulously controlled to prevent medical accidents. In this paper, an ultrasonic flowmeter (UF) with 1-3 piezoelectric composite transducers was designed, fabricated, and tested on a variety of flow rates of mimic medical injections. The transducer wedge for the angled beam propagation and an acoustic impedance matching layer were included in the design for transmission enhancement. To ensure an accurate measurement of flow rate, the effect of the flow distributions inside the IV tube was taken into account. The developed UF exhibited the capability of detecting low flow rates (<;0.005 m/s), with 1%-2% discrepancy compared with the reference rate of infusion.Keywords
This publication has 31 references indexed in Scilit:
- Errors in the administration of intravenous medications in hospital and the role of correct procedures and nurse experienceBMJ Quality & Safety, 2011
- Air–water two-phase flow measurement using a Venturi meter and an electrical resistance tomography sensorFlow Measurement and Instrumentation, 2010
- When the 5 Rights Go WrongJournal of Nursing Care Quality, 2010
- Standard Drug Concentrations and Smart-Pump Technology Reduce Continuous-Medication-Infusion Errors in Pediatric PatientsPediatrics, 2005
- Flowmetering of non-Newtonian liquidsFlow Measurement and Instrumentation, 2004
- Wet gas metering with a horizontally mounted Venturi meterFlow Measurement and Instrumentation, 2002
- Thermal flow sensor for liquids and gases based on combinations of two principlesSensors and Actuators A: Physical, 1999
- Ultrasonic metering of flow velocity based on transit time differentialsJournal of Physics E: Scientific Instruments, 1987
- Effect of volume and rate of contrast medium injection on intravenous digital subtraction angiographic contrast medium curvesJACC, 1984
- Ultrasonic flowmetersTransactions of the Institute of Measurement and Control, 1981