Direct analysis in real time ‐ high resolution mass spectrometry (DART‐HRMS): a high throughput strategy for identification and quantification of anabolic steroid esters
- 28 September 2014
- journal article
- research article
- Published by Wiley in Drug Testing and Analysis
- Vol. 7 (7) , 603-608
Abstract
High throughput screening is essential for doping, forensic, and food safety laboratories. While hyphenated chromatography‐mass spectrometry (MS) remains the approach of choice, recent ambient MS techniques, such as direct analysis in real time (DART), offer more rapid and more versatile strategies and thus gain in popularity. In this study, the potential of DART hyphenated with Orbitrap‐MS for fast identification and quantification of 21 anabolic steroid esters has been evaluated. Direct analysis in high resolution scan mode allowed steroid esters screening by accurate mass measurement (Resolution = 60 000 and mass error < 3 ppm). Steroid esters identification was further supported by collision‐induced dissociation (CID) experiments through the generation of two additional ions. Moreover, the use of labelled internal standards allowed quantitative data to be recovered based on isotopic dilution approach. Linearity (R2 > 0.99), dynamic range (from 1 to 1000 ng mL‐1), bias (‐1), repeatability and reproducibility (RSD < 20%) were evaluated as similar to those obtained with hyphenated chromatography‐mass spectrometry techniques. This innovative high throughput approach was successfully applied for the characterization of oily commercial preparations, and thus fits the needs of the competent authorities in the fight against forbidden or counterfeited substances. Copyright © 2014 John Wiley & Sons, Ltd.Keywords
This publication has 34 references indexed in Scilit:
- Anabolic—Androgenic SteroidsDisease-a-Month, 2013
- In Vitro Androgen Bioassays as a Detection Method for Designer AndrogensSensors, 2013
- Direct Peel Monitoring of Xenobiotics in Fruit by Direct Analysis in Real Time Coupled to a Linear Quadrupole Ion Trap–Orbitrap Mass SpectrometerAnalytical Chemistry, 2013
- Application of direct analysis in real time ionization–mass spectrometry (DART–MS) in chicken meat metabolomics aiming at the retrospective control of feed fraudMetabolomics, 2013
- Improvement of estradiol esters monitoring in bovine hair by dansylation and liquid chromatography/tandem mass spectrometry analysis in multiple reaction monitoring and precursor ion scan modesRapid Communications in Mass Spectrometry, 2012
- Targeted and untargeted profiling of biological fluids to screen for anabolic practices in cattleTrAC Trends in Analytical Chemistry, 2010
- Desorption electrospray ionisation mass spectrometry: A rapid screening tool for veterinary drug preparations and forensic samples from hormone crime investigationsAnalytica Chimica Acta, 2009
- A half-century of anabolic steroids in sportSteroids, 2009
- Versatile New Ion Source for the Analysis of Materials in Open Air under Ambient ConditionsAnalytical Chemistry, 2005
- Analysis of Non-Hormonal Nutritional Supplements for Anabolic-Androgenic Steroids - Results of an International StudyInternational Journal of Sports Medicine, 2004