Electromagnetic Interference Between Left Ventricular Assist Device and a Three-Dimensional Mapping System Overcome by “Hot Mapping”
- 1 July 2018
- journal article
- research article
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 64 (4) , e61-e63
Abstract
The number of patients with implanted left ventricular assist devices is constantly increasing. Numerous patients suffer from drug-resistant ventricular tachycardias. Only a few studies and reports about ablation therapy in this cohort of patients are available. Electromagnetic interferences between the left ventricular assist device and the three-dimensional mapping system which is used for ablation has been described as disabling the investigator to create a proper map of the left ventricle. We observed that this interference is interrupted by “hot mapping” meaning that one pretends an ablation with only 5 W, permitting the creation of a clear map in areas of interference.Keywords
This publication has 7 references indexed in Scilit:
- Catch 22: a case of incessant ventricular tachycardia post-left ventricular assist device resulting in right ventricular failure, left ventricular cavity obliteration and failure of endocardial ventricular tachycardia ablationJournal of Artificial Organs, 2017
- Reduction of Fluoroscopic Exposure Using a New Fluoroscopy Integrating Technology in a 3D-Mapping System During Pulmonary Vein Isolation With a Circular Multipolar Irrigated CatheterInternational Heart Journal, 2016
- Characteristics of Ventricular Tachycardia Ablation in Patients With Continuous Flow Left Ventricular Assist DevicesCirculation: Arrhythmia and Electrophysiology, 2015
- Catheter ablation of electrical storm in a patient with left ventricular assist deviceHerzschrittmachertherapie + Elektrophysiologie, 2014
- Catheter Ablation for Ventricular Tachyarrhythmias in Patients Supported by Continuous-Flow Left Ventricular Assist DevicesASAIO Journal, 2014
- Clinical implications of left ventricular assist device implantation in patients with an implantable cardioverter-defibrillatorJournal of Interventional Cardiac Electrophysiology, 2013
- Epidemiology, management, and outcomes of sustained ventricular arrhythmias after continuous-flow left ventricular assist device implantationAmerican Heart Journal, 2012