Comparison of visual and subjective outcomes between two EDOF intraocular lenses, with emphasis on mesopic performance
- 1 June 2026
- journal article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 52 (6) , 574-581
Abstract
Purpose: To evaluate outcomes after implantation of 2 extended depth-of-focus intraocular lenses (IOLs), with emphasis on performance in dim light. Setting: 2 clinical practices in the United States. Design: Prospective, double-masked, randomized clinical trial. Methods: Participants were randomized to receive Symfony Optiblue or Vivity IOLs. Objective optical quality, aberrations, and simulated visual disturbances were measured before and 3 months postoperatively. Binocular visual acuity (VA), the defocus curve, and low contrast VA were measured under mesopic lighting conditions. Patient questionnaires were administered, and undirected patient comments were recorded. Results: 31 participants in each group completed the study. Refractive outcomes were not significantly different. The Symfony lens provided about a half line better monocular mesopic VA at distance when uncorrected and best corrected ( P < .05). There were no differences in binocular visual function between lenses except for the mesopic defocus curve at −1.00 diopters (D) and −1.50 D, with the Symfony lens providing better VA at those vergences ( P < .02). Objective optical measures improved after surgery, with improvements not different by group. There were no differences in the size or intensity of simulated glare, but simulated halos and starburst size and intensity were significantly worse with the Symfony lens ( P < .05). Satisfaction with distance and intermediate vision in both groups was high. Reported visual disturbances were not significantly different between groups. Conclusions: The 2 lenses provided similar visual performance in dim light, with the Symfony lens showing slightly better VA, but with more reports of halos and starbursts.Keywords
This publication has 13 references indexed in Scilit:
- Comparison of visual performance and quality of life with a new nondiffractive EDOF intraocular lens and a trifocal intraocular lensJournal of Cataract & Refractive Surgery, 2023
- Effect of Violet Light-Filtering and Manufacturing Improvements in an Extended Depth-of-Focus Intraocular Lens on Visual PerformanceClinical Ophthalmology, 2023
- Clinical outcomes in a U.S. registration study of a new EDOF intraocular lens with a nondiffractive designJournal of Cataract & Refractive Surgery, 2022
- Visual outcomes and safety of an extended depth-of-focus intraocular lens: results of a pivotal clinical trialJournal of Cataract & Refractive Surgery, 2021
- Multicountry clinical outcomes of a new nondiffractive presbyopia-correcting IOLJournal of Cataract & Refractive Surgery, 2021
- Comparison of visual outcomes after implantation of AtLisa tri 839 MP and Symfony intraocular lensesInternational Ophthalmology, 2020
- Visual performance comparison of 2 extended depth-of-focus intraocular lensesJournal of Cataract & Refractive Surgery, 2020
- Extended depth-of-focus technology in intraocular lensesJournal of Cataract & Refractive Surgery, 2020
- Visual performance of an extended depth of focus intraocular lens for treatment selectionEye, 2019
- Patient-Reported Spectacle Independence Questionnaire (PRSIQ): Development and ValidationAmerican Journal of Ophthalmology, 2017