Evaluation of sleeved implant fixation using a rat model
Authors:
- Yong-San YoonYYYong-San Yoon
- Deuksoo HwangDHDeuksoo Hwang
- 18 November 2010
- journal article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 33 (3) , 310-314
Abstract
Cementless hip implants have some osteolysis, stress shielding, and/or revision problems while the cemented ones have cement fracture problems. To overcome these problems, a new concept of implant fixation using a sleeve between the implant and the bone is proposed. This is a cementless fixation using a sleeve in place of the cement with a press-fit stem. The sleeved implant fixation is compared with the conventional cemented fixation by measuring the bone volume fraction and interface strength in a rat model. The tapered implant was manufactured using a stainless-steel Kirschner-wire with a diameter of 1.6mm, and PEEK (polyetheretherketone) was used for the sleeve. The Sprague-Dawley rats had implantations in their bilateral diaphyseal cut femurs using cemented and sleeved fixation, respectively. The rats were sacrificed at 2, 4, and 12 weeks after the implantation, and micro-CT scanning as well as push-out testing were performed. There was no significant difference between the cemented fixation and the sleeved fixation in areas of both BV/TV (bone volume/total volume) (p=0.838) and the apparent interface strength (p=0.530). However, the decrease in the BV/TV over time was significant in the cemented fixation (p=0.021) while not in the sleeved fixation (p=0.603). The sleeved fixation showed a gain in strength with time while the cemented fixation showed a reverse tendency but the difference was not significant (p=0.125). The results showed the feasibility of the sleeved fixation even though the advantage of sleeved fixation was not clearly confirmed because of the small sample size and high variance as well as the suboptimal surface finish of the sleeve.Keywords
This publication has 34 references indexed in Scilit:
- Changes in periprosthetic bone remodelling after redesigning an anatomic cementless stemInternational Orthopaedics, 2008
- Uncemented short‐length diaphyseal segmental replacement prosthesis fixation—Finite element analysis and long‐term resultsJournal of Orthopaedic Research, 2005
- Saline irrigation does not affect bone formation or fixation strength of hydroxyapatite/tricalcium phosphate‐coated implants in a rat modelJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2005
- A low‐temperature biomimetic calcium phosphate surface enhances early implant fixation in a rat modelJournal of Biomedical Materials Research Part A, 2004
- Survival analysis of cementless grit-blasted titanium total hip arthroplastiesThe Journal of Bone and Joint Surgery. British volume, 2001
- Influence of stem geometry on mechanics of cemented femoral hip components with a proximal bondJournal of Orthopaedic Research, 1997
- Cancellous bone stresses surrounding the femoral component of a hip prosthesis: an elastic-plastic finite element analysisMedical Engineering & Physics, 1995
- Effects of Fit and Bonding Characteristics of Femoral Stems on Adaptive Bone RemodelingJournal of Biomechanical Engineering, 1994
- The mechanism of bone remodeling and resorption around press-fitted THA stemsJournal of Biomechanics, 1993
- Intermittent micromotion inhibits bone ingrowthActa Orthopaedica, 1992