Comparison of mid-term survival of uncemented total knee arthroplasties with modular tibial components made of porous tantalum vs. cemented total knee arthroplasties
- 6 October 2025
- journal article
- Published by Taylor & Francis in Expert Review of Medical Devices
- Vol. 22 (12) , 1415-1423
Abstract
Cementless total knee arthroplasties (TKAs) with porous tantalum tibial components were developed to improve fixation and reduce implant failure rates in younger, active patients. However, mid-term outcomes remain uncertain. This retrospective study compared survival rates of cementless TKAs with porous tantalum posterior-stabilized (PS) modular tibial components to cemented PS components from the same implant system, using revision for aseptic loosening as the endpoint. From January 2017 to November 2021, 899 TKAs were performed at a single tertiary center for osteoarthritis. Of these, 645 (71.7%) were cemented and 254 (28.3%) cementless. Median BMI was similar in both groups (30.0 kg/m2 vs 31.0 kg/m2; p = 0.814). Cementless recipients were younger (median 66.0 vs. 72.0 years; p < 0.001). Follow-up continued through 31 December 2023. Seven-year survival was lower in the cementless group (97.1%, 95% CI 94.9-99.3) vs. cemented group (97.8%, 95% CI 96.6-99.0). Tibial component loosening led to 5 revisions in the cementless group vs. 2 revisions in the cemented group (p < 0.05). Two cementless cases involved tibial tray fractures. Overall, 86 patients (9.6%) died of unrelated causes. Cementless porous tantalum TKAs had inferior mid-term performance for aseptic loosening, prompting their discontinuation at our institution.Keywords
This publication has 41 references indexed in Scilit:
- Tantalum Particles Promote M2 Macrophage Polarization and Regulate Local Bone Metabolism via Macrophage‐Derived Exosomes Influencing the Fates of BMSCsAdvanced Healthcare Materials, 2024
- Trabecular metal monoblock versus modular tibial trays in total knee arthroplasty: meta-analysis of randomized control trialsInternational Orthopaedics, 2022
- A comprehensive review on metallic implant biomaterials and their subtractive manufacturingThe International Journal of Advanced Manufacturing Technology, 2022
- Catastrophic tibial baseplate failure of a modern cementless total knee arthroplasty implantArthroplasty Today, 2019
- Assessment of thermal damage in total knee arthroplasty using an osteocyte injury modelJournal of Orthopaedic Research, 2017
- The Mark Coventry AwardClinical Orthopaedics and Related Research, 2015
- Trabecular Metal in Total Knee Arthroplasty Associated with Higher Knee Scores: A Randomized Controlled TrialClinical Orthopaedics and Related Research, 2013
- Do Porous Tantalum Implants Help Preserve Bone?: Evaluation of Tibial Bone Density Surrounding Tantalum Tibial Implants in TKAClinical Orthopaedics and Related Research, 2010
- Preliminary Results of an Uncemented Trabecular Metal Tibial Component in Total Knee ArthroplastyThe Journal of Arthroplasty, 2009
- Porous tantalum in hip and knee reconstructive surgeryJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2008