Carbon Dioxide Absorbents Containing Potassium Hydroxide Produce Much Larger Concentrations of Compound A from Sevoflurane in Clinical Practice
Authors:
- Michiaki YamakageMYMichiaki Yamakage
- Shigeyuki YamadaSYShigeyuki Yamada
- Xiangdong ChenXCXiangdong Chen
- Sohshi IwasakiSISohshi Iwasaki
- Naoki TsujiguchiNTNaoki Tsujiguchi
- Akiyoshi NamikiANAkiyoshi Namiki
- 1 July 2000
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 91 (1) , 220-224
Abstract
We investigated the concentrations of degraded sevoflurane Compound A during low-flow anesthesia with four carbon dioxide (CO(2)) absorbents. The concentrations of Compound A, obtained from the inspiratory limb of the circle system, were measured by using a gas chromatograph. In the groups administered 2 L/min fresh gas flow with 1% sevoflurane, when the conventional CO(2) absorbents, Wakolime(TM) (Wako, Tokyo, Japan) and Drägersorb(TM) (Dräger, Lübeck, Germany), were used, the concentrations of Compound A increased steadily from a baseline to 14.3 ppm (mean) and 13.2 ppm, respectively, at 2 h after exposure to sevoflurane. In contrast, when the other novel types of absorbents containing decreased or no potassium hydroxide/sodium hydroxide, Medisorb(TM) (Datex-Ohmeda, Louisville, CO) and Amsorb(TM) (Armstrong, Coleraine, Northern Ireland), were used, Compound A remained at baseline (<2 ppm) throughout the study. In the groups administered 1 L/min fresh gas flow with 2% sevoflurane, Wakolime(TM) and Drägersorb(TM) produced much larger concentrations of Compound A (35.4 ppm and 34.2 ppm, respectively) at 2 h after exposure to sevoflurane. Medisorb(TM) showed measurable concentrations of Compound A (8.6 ppm at 2 h), but they were significantly smaller than those produced by the two conventional absorbents. In contrast, when Amsorb(TM) was used, Compound A concentrations remained at baseline throughout the study period. Carbon dioxide absorbents containing potassium hydroxide/sodium hydroxide produce much larger concentrations of Compound A from sevoflurane in clinical practice. An absorbent containing neither potassium hydroxide nor sodium hydroxide produces the smallest concentrations of Compound A.Keywords
This publication has 20 references indexed in Scilit:
- AmsorbAnesthesiology, 1999
- Dose of Compound A, Not Sevoflurane, Determines Changes in the Biochemical Markers of Renal Injury in Healthy VolunteersAnesthesia & Analgesia, 1999
- A Case of Nonoliguric Renal Failure After General Anesthesia with Sevoflurane and DesfluraneAnesthesia & Analgesia, 1997
- Effects of Low-flow Sevoflurane Anesthesia on Renal FunctionAnesthesiology, 1997
- Low-flow (1 l/min) SevofluraneAnesthesiology, 1997
- Nephrotoxicity of Sevoflurane Versus Desflurane Anesthesia in VolunteersAnesthesia & Analgesia, 1997
- Inhalation Toxicity Study of a Haloalkene Degradant of Sevoflurane, Compound A (PIFE), in Sprague-Dawley RatsAnesthesiology, 1995
- Long-duration, Low-flow Sevoflurane Anesthesia Using Two Carbon Dioxide AbsorbentsAnesthesiology, 1994
- Toxicity of Compound A in RatsAnesthesiology, 1994
- SevofluraneAnesthesia & Analgesia, 1975