Validation and Precision of Model-Based Radiostereometric Analysis (MBRSA) for Total Ankle Arthroplasty
- 1 December 2011
- journal article
- research article
- Published by SAGE Publications in Foot & Ankle International
- Vol. 32 (12) , 1155-1163
Abstract
Background: The goal of this study was to design a RSA marker insertion protocol to evaluate the stability of the bone-implant interface of a TAA prosthesis, and to validate that this marker insertion protocol can be combined with MBRSA technology to provide clinically adequate precision in assessing the micromotion of the TAA prosthesis. Methods: The Mobility™ Total Ankle System was used in this study. A marker placement protocol was developed with a Phantom Protocol. The Improved Marker Placement Protocol was used in 20 patients. Postoperative RSA double exams were taken. Condition Numbers (CN) were used to assess the marker distribution. The system precision was defined as the standard deviation of the double exams (MTE, MRE). MBRSA software was used to evaluate the double exams. Results: The RSA marker insertion technique for the 20 in vivo cases provided satisfactory results. CNs in all subjects but one were below 50 mm−1 and implied a desirable marker configuration. The tibial sphere MTE was 0.07 mm and the talar was 0.09 mm. The talar MRE was 0.51 degrees. Conclusion: The system precision for these in vivo TAA implants was within the normal range identified by RSA studies, and comparable to the existing TAA RSA studies. This study demonstrated a reliable RSA marker insertion technique in both the tibia and talus. The study confirms that the insertion and MBRSA technique allows the typical high precision demonstrated in other RSA studiesKeywords
This publication has 36 references indexed in Scilit:
- Implementation and validation of an implant-based coordinate system for RSA migration calculationJournal of Biomechanics, 2009
- Experimental Analysis of Model-Based Roentgen Stereophotogrammetric Analysis (MBRSA) on Four Typical Prosthesis ComponentsJournal of Biomechanical Engineering, 2009
- History and evolution in total ankle arthroplastyBritish Medical Bulletin, 2008
- The History and Future of Radiostereometric AnalysisClinical Orthopaedics and Related Research, 2006
- Model-based RSA of a Femoral Hip Stem Using Surface and Geometrical Shape ModelsClinical Orthopaedics and Related Research, 2006
- Early Migration of the Tibial Component of the Buechel-Pappas Total Ankle ProsthesisClinical Orthopaedics and Related Research, 2006
- Guidelines for standardization of radiostereometry (RSA) of implantsActa Orthopaedica, 2005
- Accuracy and precision of radiostereometric analysis in the measurement of THR femoral component translations: human and canine in vitro modelsJournal of Orthopaedic Research, 2001
- Methods for determining the accuracy of radiostereometric analysis (RSA)Acta Orthopaedica, 2000
- Determining the movements of the skeleton using well-configured markersJournal of Biomechanics, 1993