No corrosion of 304 stainless steel implant after 40 years of service
Authors:
- 1 July 2004
- journal article
- Published by Springer Nature in Journal of Materials Science: Materials in Medicine
- Vol. 15 (7) , 755-758
Abstract
When exposed to 0.9% NaCl type 304 stainless steel undergoes severe pitting corrosion within a matter of days. However, a Sherman plate fabricated from type 304 stainless steel remained inside a...Keywords
This publication has 18 references indexed in Scilit:
- The influence of complexing agent and proteins on the corrosion of stainless steels and their metal componentsJournal of Materials Science: Materials in Medicine, 2003
- Evolution of the internal fixation of long bone fractures: The scientific basis of biological internal fixation: choosing a new balance between stability and biologyThe Journal of Bone and Joint Surgery. British volume, 2002
- The behavior of stainless steels in physiological solution containing complexing agent studied by X-ray photoelectron spectroscopyJournal of Biomedical Materials Research, 2000
- Temporal Expression of the Chondrogenic and Angiogenic Growth Factor CYR61 During Fracture RepairJournal of Bone and Mineral Research, 2000
- Electrochemical and surface modifications on N+-ion-implanted 316 L stainless steelJournal of Materials Science: Materials in Medicine, 1997
- An in-vitro study of H2O2-treated titanium surfaces in contact with blood plasma and a simulated body fluidClinical Materials, 1993
- Physico-chemical considerations of titanium as a biomaterialClinical Materials, 1992
- Schadensfälle an chirurgischen Implantaten und ihre UrsachenMaterials and Corrosion, 1991
- Schadensfälle an chirurgischen Implantaten und deren UrsachenMaterials and Corrosion, 1981
- [Corrosion of a Sherman plate implanted on the humerus over a 48-year period. Clinical and microscopic study].1974