Use of a multi‐electrode radiofrequency balloon catheter to achieve pulmonary vein isolation in patients with paroxysmal atrial fibrillation: 12‐Month outcomes of the RADIANCE study
- 6 April 2020
- journal article
- research article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 31 (6) , 1259-1269
Abstract
Background: The RADIANCE first‐in‐man study evaluated acute (3‐month) safety and design concept in terms of utility of a new multi‐electrode radiofrequency (RF) balloon catheter (HELIOSTAR, Biosense Webster) to achieve pulmonary vein isolation (PVI). After study conclusion, a subset of patients was followed up to 12 months.Methods: Patients with drug refractory paroxysmal atrial fibrillation were enrolled. Neurological assessment, cardiac and cerebral magnetic resonance imagings were performed pre and post procedure. Ablation was delivered at 15 Watts to each PV for 60 seconds (electrodes adjacent to the posterior wall limited to 20 seconds). Adenosine or isoproterenol was administered to confirm PVI. Esophageal endoscopy was performed 48 hours post procedure. Patients were clinically followed up for 12 months.Results: Thirty‐nine patients underwent catheter ablation from four centers. Mean age was 60.7 ± 10.0 years with 23 (57.5%) being male. Confirmation of PVI was performed in all PVs treated (152/152). Confirmation of isolation after one delivery was performed solely on 137 of 152 PVs of which 79.6% (109/137) achieved isolation with a single delivery of RF energy. Acute PV reconnection was seen in 4.6% (7/150) of PVs. Freedom from documented atrial arrhythmia at 12 months in those followed up was 86.4% (32/37). A total of 75.7% (28/37) of patients were free from atrial arrhythmia and off antiarrhythmic medications.Conclusion: The HELIOSTAR RF balloon catheter allows for rapid and safe PVI with majority of PVs only requiring one application.Keywords
Funding Information
- Biosense Webster
This publication has 23 references indexed in Scilit:
- Impact of Complete Versus Incomplete Circumferential Lines Around the Pulmonary Veins During Catheter Ablation of Paroxysmal Atrial FibrillationCirculation: Arrhythmia and Electrophysiology, 2016
- First-line catheter ablation of paroxysmal atrial fibrillation: outcome of radiofrequency vs. cryoballoon pulmonary vein isolationEP Europace, 2015
- Adenosine-guided pulmonary vein isolation for the treatment of paroxysmal atrial fibrillation: an international, multicentre, randomised superiority trialThe Lancet, 2015
- Adenosine testing after second-generation cryoballoon ablation (ATSCA) study improves clinical success rate for atrial fibrillationEP Europace, 2015
- Spontaneous and Adenosine‐Induced Pulmonary Vein Reconnection After Cryoballoon Ablation with the Second‐Generation DeviceJournal of Cardiovascular Electrophysiology, 2014
- Updated Worldwide Survey on the Methods, Efficacy, and Safety of Catheter Ablation for Human Atrial FibrillationCirculation: Arrhythmia and Electrophysiology, 2010
- Acute and Chronic Pulmonary Vein Reconnection after Atrial Fibrillation Ablation: A Prospective Characterization of Anatomical SitesPacing and Clinical Electrophysiology, 2008
- Anatomic Relations Between the Esophagus and Left Atrium and Relevance for Ablation of Atrial FibrillationCirculation, 2005
- Perception of Atrial Fibrillation Before and After Radiofrequency Catheter AblationCirculation, 2005
- How Close Are the Phrenic Nerves to Cardiac Structures? Implications for Cardiac InterventionalistsJournal of Cardiovascular Electrophysiology, 2005