Incidence of lead dislodgement, malfunction and perforation during the first year following device implantation
Authors:
Open Access
- 8 May 2014
- journal article
- research article
- Published by Springer Nature in Netherlands Heart Journal
- Vol. 22 (6) , 286-291
Abstract
Background The number of cardiac rhythm device implantations has been growing fast due to expanding indications and ageing of the population. Complications of implantation were rare in the trials. However, these involved small numbers and selected patients. Prospective real-life data are necessary to assess cardiac device implantation procedure-related risks. Objective To determine the incidence and predictors of lead-related re-intervention in a Dutch high-volume teaching hospital. Methods Data from all patients who underwent cardiac rhythm device implantation between January 2010 and December 2011 were collected in a prospective registry. At least 1 year of follow-up regarding re-intervention was available for all patients. Lead-related reasons for re-intervention were categorised into lead dislodgement, malfunctioning or perforation. Results One thousand nine hundred twenty-nine devices including 3909 leads were implanted. In 595 patients (30.8 %) a CRT-D/P was implanted. Lead-related re-intervention was necessary in 86 (4.4 %) patients; it was more common in younger and male patients, and due to either lead dislodgement (66 %), malfunctioning (20 %) or perforation (18 %). Coronary sinus lead dislodgement or malfunctioning was 1.4 %. Right atrial dislodgement (1.9 %, p < 0.001) or ICD lead dislodgement (1.8 %, p = 0.002) was more common than right ventricular dislodgement (0.3 %). The incidence of lead malfunctioning was higher (0.8 %) in ICD leads. An apical position of the right ventricular lead and lateral wall position of the right atrial lead were related to cardiac perforation. Conclusions The incidence of lead-related re-intervention was comparable with the literature. The majority of re-interventions were due to lead dislodgements, particularly with right atrial and ICD leads. Re-intervention due to coronary sinus lead dislodgement was rare.Keywords
This publication has 18 references indexed in Scilit:
- Rate, causes, and impact on patient outcome of implantable device complications requiring surgical revision: large population survey from two centres in ItalyEP Europace, 2013
- Are Elderly Patients at Increased Risk of Complications Following Pacemaker Implantation? A Meta‐Analysis of Randomized TrialsPacing and Clinical Electrophysiology, 2011
- Implantation-Related Complications of Implantable Cardioverter-Defibrillators and Cardiac Resynchronization Therapy DevicesJACC, 2011
- Acute Lead Dislodgements and In-Hospital Mortality in Patients Enrolled in the National Cardiovascular Data Registry Implantable Cardioverter Defibrillator RegistryJACC, 2010
- Cardiac-Resynchronization Therapy for the Prevention of Heart-Failure EventsNew England Journal of Medicine, 2009
- Subacute cardiac perforations associated with active fixation leadsEP Europace, 2008
- Guidelines for cardiac pacing and cardiac resynchronization therapy: The Task Force for Cardiac Pacing and Cardiac Resynchronization Therapy of the European Society of Cardiology. Developed in Collaboration with the European Heart Rhythm AssociationEP Europace, 2007
- Implantation of cardiac resynchronization therapy systems in the CARE-HF trial: procedural success rate and safetyEP Europace, 2007
- Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillatorHeart Rhythm, 2007
- Primary Prevention of Sudden Cardiac Death in Idiopathic Dilated CardiomyopathyCirculation, 2002