In ultrasound B-mode imaging, Filtered Delay Multiply and Sum (F-DMAS) could gain a higher contrast resolution for anechoic cysts (a narrower mainlobe and lower sidelobes) and better noise rejection than the traditional Delay and Sum (DAS) beamforming method. However, a phase-delay distortion may be brought into the reconstructed image by the bandpass filters after Delay Multiply and Sum (F-DMAS) beamforming. In this paper, a zero-phase bandpass filter was adopted to eliminate the phase-delay distortion. The performance of the proposed method was evaluated by the developed ultrasound computed tomography with breast phantom data. Experimental results show that the zero-phase filtered delay multiply and sum (ZPF-DMAS) beamforming can reduce the sidelobes or grating lobes and restrain the noises and artifacts without phase-delay distortion. The image contrast ratio can be enhanced about 0.4 dB and 4.6 dB respectively for cystic and dense masses and contrast to noise ratio about 0.04 dB and 0.5 dB respectively for cystic and dense masses, compared to the traditional DAS beamforming algorithm.