Continuous glucose monitoring following pancreatectomy with islet autotransplantation in children

Authors:

Abstract
Aim was to determine whether CGM could accurately monitor blood glucose concentration in the immediate postoperative period following pancreatectomy with IAT in children. CGM was used in nine patients undergoing IAT at our institution between April 2015 and September 2016 (eight total pancreatectomy and one subtotal pancreatectomy). MAD and MARD of CGM values compared to time‐matched serum blood glucose were calculated during the first 5 days of ICU admission. Goal range was defined as 70‐140 mg/dL and out‐of‐range was >140 mg/dL or P=.74). By Clarke error grid analysis, 100% of paired values were in clinically acceptable zones. By surveillance error grid analysis, 96% of paired values were within clinically acceptable agreement. CGM is a reliable tool in monitoring glycemic control in the immediate postoperative period following pancreatectomy with IAT in children.
Funding Information
  • Cincinnati Children's Hospital Medical Center Academic and Research Committee

This publication has 18 references indexed in Scilit: