Validation of the visocor HM40 wrist blood pressure measuring device according to the International Protocol
Authors:
- Elisa BonsoEBElisa Bonso
- Francesca SaladiniFSFrancesca Saladini
- Paolo PalatiniPPPaolo Palatini
- 1 April 2009
- journal article
- research article
- Published by Wolters Kluwer Health in Blood Pressure Monitoring
- Vol. 14 (2) , 83-86
Abstract
The objective of this study was to determine the accuracy of the UEBE visocor HM40 device for blood pressure measurement at the wrist, tested according to the requirements of the International Protocol of the European Society of Hypertension (ESH). Device evaluation was performed in 33 patients whose age and blood pressure were within the intervals required by the ESH protocol. Their mean±SD age was 53.2±15.5 years (range 30–85 years), systolic blood pressure (SBP) was 142.8±21.8 mmHg (range 108–178 mmHg), diastolic blood pressure (DBP) was 87.0±15.4 mmHg (range 62–112 mmHg), and wrist circumference was 18.7±2.7 cm (range 15.5–21.5 cm). During the measurements, which were performed in the supine position, the patient's wrist was kept extended at the heart level with a support under the forearm. In total, 45 measurements were available for analysis in the first phase of the validation process, and 99 measurements in the second phase. The visocor HM40 passed all three phases of the ESH protocol for SBP and DBP. Mean blood pressure differences for the visocor HM40 monitor (device–observer) were −2.0±5.0 mmHg for SBP and −1.7±5.3 mmHg for DBP, respectively. These data show that the UEBE visocor HM40 wrist monitor met the requirements for validation by the International Protocol and can be recommended for clinical use in the adult population.Keywords
This publication has 15 references indexed in Scilit:
- Trial of validation of two devices for self-measurement of blood pressure according to the European Society of Hypertension International Protocol: the Citizen CH-432B and the Citizen CH-656CBlood Pressure Monitoring, 2008
- Validation of the Samsung SBM-100A and Microlife BP 3BU1-5 wrist blood pressure measuring devices in adults according to the International ProtocolBlood Pressure Monitoring, 2007
- Rapid Assessment of Blood Pressure in the Obstetric Day Unit Using Microlife MaM TechnologyHypertension in Pregnancy, 2007
- Inaccuracy of wrist-cuff oscillometric blood pressure devices: an arm position artefact?Blood Pressure Monitoring, 2005
- What will replace the mercury sphygmomanometer?Blood Pressure Monitoring, 2003
- Accuracy and reliability of wrist-cuff devices for self-measurement of blood pressureJournal Of Hypertension, 2002
- Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adultsBlood Pressure Monitoring, 2002
- Has conventional sphygmomanometry ended with the banning of mercury?Blood Pressure Monitoring, 2002
- Oscillometric blood pressure measurement: progress and problemsBlood Pressure Monitoring, 2001
- Reporting bias in self-measurement of blood pressureBlood Pressure Monitoring, 2001