Functionality of implanted microchips following magnetic resonance imaging
Authors:
- Katherine A. HaifleyKHKatherine A. Haifley
- Silke HechtSHSilke Hecht
- 1 March 2012
- journal article
- Published by American Veterinary Medical Association (AVMA) in Journal of the American Veterinary Medical Association
- Vol. 240 (5) , 577-579
Abstract
Objective—To determine the functionality of implanted microchips following magnetic resonance imaging (MRI). Design—Prospective clinical trial. Animals—53 client-owned patients implanted with a microchip undergoing MRI of various areas of the body for a variety of medical conditions. Procedures—General anesthesia was induced, and each patient's microchip was scanned with a universal microchip scanner; the chip number was recorded. Patients were transported to the MRI suite, and MRI was completed. Patients were moved out of the magnetic environment, and microchips were scanned again. Patient information and chip number were recorded. Chip numbers before and after MRI were compared. Results—For all 53 microchips scanned from 53 patients, the same number was read accurately following MRI of a variety of sites. Conclusions and Clinical Relevance—These data indicated that MRI did not interfere with the functionality of these microchips. This information is valuable for practitioners recommending MRI for their patients and for clients who have invested in implanting a microchip in their pets.Keywords
This publication has 5 references indexed in Scilit:
- Evaluation of the Susceptibility Artifacts and Tissue Injury Caused by Implanted Microchips in Dogs on 1.5T Magnetic Resonance ImagingJournal of Veterinary Medical Science, 2010
- Characterization of animals with microchips entering animal sheltersJournal of the American Veterinary Medical Association, 2009
- Sensitivity of commercial scanners to microchips of various frequencies implanted in dogs and catsJournal of the American Veterinary Medical Association, 2008
- Search and identification methods that owners use to find a lost dogJournal of the American Veterinary Medical Association, 2007
- MRI Physics for Physicians1989