Abstract
Absolute distance measurements can be performed with an interferometric method that uses only a single tunable laser. This method has one major drawback, because a small target movement of the order of one wavelength during a measurement will be interpreted as a movement of one synthetic wavelength. This effect is usually mitigated by adding a second (nonscanning) laser. We show that absolute distance measurements can be performed with only one laser if the movements encountered are smooth, on the time scale of one measurement. In this case the movement errors can be compensated with a simple algorithm that combines several subsequent measurements. First experimental results show good agreement with theory.Keywords
This publication has 7 references indexed in Scilit:
- Absolute distance metrology for space interferometers2004
- MSTAR: a submicrometer absolute metrology systemOptics Letters, 2003
- Reducing the linewidth of a diode laser below 30 Hz by stabilization to a reference cavity with a finesse above 10^5Optics Letters, 2001
- High-accuracy absolute distance measurement by means of wavelength scanning heterodyne interferometryMeasurement Science and Technology, 1998
- Absolute interferometric distance measurements applying a variable synthetic wavelengthJournal of Optics, 1998
- Interferometric measurement of absolute distances of up to 40 mMeasurement, 1995
- Rangefinder with fast multiple range capabilityReview of Scientific Instruments, 1992