Comparison of the Multiple Reaction Monitoring and Enhanced Product Ion Scan Modes for Confirmation of Stilbenes in Bovine Urine Samples Using LC–MS/MS QTRAP® System
Open Access
- 5 July 2016
- journal article
- research article
- Published by Springer Nature in Chromatographia
- Vol. 79 (15-16) , 1003-1012
Abstract
In accordance with Commission Decision 2002/657/EC, confirmatory methods for the detection of prohibited substances should comply with specific requirements, including the criteria for confirmation. Two strategies: multiple reaction monitoring (MRM) and enhanced product ion (EPI) scanning functions were compared for confirming the anabolic compounds from synthetic stilbenes group in bovine urine samples. In the research, twenty samples fortified at the Recommended Concentration (RC) of 1 µg L−1 with diethylstilbestrol, dienestrol and hexestrol were analyzed by liquid chromatography-tandem mass spectrometry on a QTRAP 5500 instrument. The analytical procedure, validated in accordance with the Commission Decision 2002/657/EC, used in the official control of hormones in Poland was applied. The validation parameters were in agreement with 2002/657/EC performance criteria. The effectiveness of MRM and EPI scanning modes for confirmation purposes was evaluated based on the percentage of the results confirmed. In all urine samples recorded in the MRM mode, the confirmation criteria (retention time, relative intensities between transitions) have been fulfilled. The presence of stilbenes in all urine samples using EPI scan mode was confirmed too as evidenced by a good matching of stilbenes spectra in the samples to the reference spectra with critical match factor above 0.7. The results of the research show that EPI scanning function provides the same effectiveness for confirmation of banned compounds as the mostly used MRM scan mode and can be an additional tool to confirm the doubtful case results in the analysis of hormones residues, even at such low concentration levels.Keywords
This publication has 22 references indexed in Scilit:
- Protocol for an Electrospray Ionization Tandem Mass Spectral Product Ion Library: Development and Application for Identification of 240 Pesticides in FoodsAnalytical Chemistry, 2012
- Routine approach to qualitatively screening 300 pesticides and quantification of those frequently detected in fruit and vegetables using liquid chromatography tandem mass spectrometry (LC-MS/MS)Food Additives & Contaminants: Part A, 2010
- Rapid confirmatory method for the determination of sixteen synthetic growth promoters and bisphenol A in bovine milk using dispersive solid-phase extraction and liquid chromatography–tandem mass spectrometryJournal of Chromatography B, 2010
- Determination of anabolic steroids in bovine urine by liquid chromatography–tandem mass spectrometryJournal of Chromatography B, 2009
- Ultra‐performance liquid chromatography/tandem mass spectrometry in high‐throughput detection, quantification and confirmation of anabolic steroids in equine plasmaRapid Communications in Mass Spectrometry, 2009
- Determination of synthetic drugs used to adulterate botanical dietary supplements using QTRAP LC-MS/MSFood Additives & Contaminants: Part A, 2009
- Simultaneous determination of residual hormonal chemicals in meat, kidney, liver tissues and milk by liquid chromatography–tandem mass spectrometryAnalytica Chimica Acta, 2005
- Development of a multi‐target screening analysis for 301 drugs using a QTrap liquid chromatography/tandem mass spectrometry system and automated library searchingRapid Communications in Mass Spectrometry, 2005
- Combination of liquid chromatography/tandem mass spectrometry and gas chromatography/mass spectrometry for the detection of 21 anabolic steroid residues in bovine urineJournal of Mass Spectrometry, 2005
- Findings in female offspring of women exposed in utero to diethylstilbestrolObstetrics & Gynecology, 2002