Variability in EIT Images of Lung Ventilation as a Function of Electrode Planes and Body Positions
Authors:
Open Access
- 27 June 2014
- journal article
- Published by Bentham Science Publishers Ltd. in The Open Biomedical Engineering Journal
- Vol. 8 (1) , 35-41
Abstract
This study is aimed at investigating the variability in resistivity changes in the lung region as a function of air volume, electrode plane and body position. Six normal subjects (33.8 ± 4.7 years, range from 26 to 37 years) were studied using the Sheffield Electrical Impedance Tomography (EIT) portable system. Three transverse planes at the level of second intercostal space, the level of the xiphisternal joint, and midway between upper and lower locations were chosen for measurements. For each plane, sixteen electrodes were uniformly positioned around the thorax. Data were collected with the breath held at end expiration and after inspiring 0.5, 1.0, or 1.5 liters of air from end expiration, with the subject in both the supine and sitting position. The average resistivity change in five regions, two 8x8 pixel local regions in the right lung, entire right, entire left and total lung regions, were calculated. The results show the resistivity change averaged over electrode positions and subject positions was 7-9% per liter of air, with a slightly larger resistivity change of 10 % per liter air in the lower electrode plane. There was no significant difference (p>0.05) between supine and sitting. The two 8x8 regions show a larger inter individual variability (coefficient of variation, CV, is from 30% to 382%) compared to the entire left, entire right and total lung (CV is from 11% to 51%). The results for the global regions are more consistent. The large inter individual variability appears to be a problem for clinical applications of EIT, such as regional ventilation. The variability may be mitigated by choosing appropriate electrode plane, body position and region of interest for the analysis.Keywords
This publication has 20 references indexed in Scilit:
- Whither lung EIT: Where are we, where do we want to go and what do we need to get there?Physiological Measurement, 2012
- Adjacent stimulation and measurement patterns considered harmfulPhysiological Measurement, 2011
- GREIT: a unified approach to 2D linear EIT reconstruction of lung imagesPhysiological Measurement, 2009
- Accounting for hardware imperfections in EIT image reconstruction algorithmsPhysiological Measurement, 2007
- EIT images of ventilation: what contributes to the resistivity changes?Physiological Measurement, 2005
- Thoracic electrical impedance tomographic measurements during volume controlled ventilation-effects of tidal volume and positive end-expiratory pressureIEEE Transactions on Medical Imaging, 1999
- Application of electrical impedance tomography in diagnosis of emphysema-a clinical studyPhysiological Measurement, 1995
- Applied potential tomography: a new technique for monitoring pulmonary functionClinical Physics and Physiological Measurement, 1988
- Applications of applied potential tomography (APT) in respiratory medicineClinical Physics and Physiological Measurement, 1987
- Fast reconstruction of resistance imagesClinical Physics and Physiological Measurement, 1987