Study of compatibility between a 3T MR system and detector modules for a second‐generation RF‐penetrable TOF‐PET insert for simultaneous PET/MRI
Authors:
- 7 April 2023
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 50 (6) , 3389-3400
Abstract
Background: Simultaneous positron emission tomography/magnetic resonance imaging (PET/MRI) has shown promise in acquiring complementary multiparametric information of disease. However, designing these hybrid imaging systems is challenging due to the propensity for mutual interference between the PET and MRI subsystems. Currently, there are integrated PET/MRI systems for clinical applications. For neurologic imaging, a brain‐dedicated PET insert provides superior spatial resolution and sensitivity compared to body PET scanners.Purpose: Our first‐generation prototype brain PET insert (“PETcoil”) demonstrated RF‐penetrability and MR‐compatibility. In the second‐generation PETcoil system, all analog silicon photomultiplier (SiPM) signal digitization is moved inside the detectors, which results in substantially better PET detector performance, but presents a greater technical challenge for achieving MR‐compatibility. In this paper, we report results from MR‐compatibility studies of two fully assembled second‐generation PET insert detector modules.Methods: We studied the effect of the presence of the two second‐generation TOF‐PET insert detectors on parameters that affect MR image quality and evaluated TOF‐PET detector performance under different MRI pulse sequence conditions.Results: With the presence of operating PET detectors, no RF noise peaks were induced in the MR images, but the relative average noise level was increased by 15%, which led to a 3.1 to 4.2‐dB degradation in MR image signal‐to‐noise ratio (SNR). The relative homogeneity of MR images degraded by less than 1.5% with the two operating TOF‐PET detectors present. The reported results also indicated that ghosting artifacts (percent signal ghosting (PSG) ⩽ 1%) and MR susceptibility artifacts (0.044 ppm) were insignificant. The PET detector data showed a relative change of less than 5% in detector module performance between running outside and within the MR bore under different MRI pulse sequences except for energy resolution in EPI sequence (13% relative difference).Conclusions: The PET detector operation did not cause any significant artifacts in MR images and the performance and time‐of‐flight (TOF) capability of the former were preserved under different tested MR conditions.Keywords
Funding Information
- National Institutes of Health (R01EB01946504)
This publication has 26 references indexed in Scilit:
- Simultaneous PET/MR imaging with a radio frequency‐penetrable PET insertMedical Physics, 2016
- NEMA NU 2-2012 performance studies for the SiPM-based ToF-PET component of the GE SIGNA PET/MR systemMedical Physics, 2016
- Performance characterization of compressed sensing positron emission tomography detectors and data acquisition systemPhysics in Medicine & Biology, 2015
- Simultaneous PET/MR head–neck cancer imaging: Preliminary clinical experience and multiparametric evaluationEuropean Journal of Radiology, 2015
- Prototype positron emission tomography insert with electro-optical signal transmission for simultaneous operation with MRIPhysics in Medicine & Biology, 2015
- Performance Measurements of the Siemens mMR Integrated Whole-Body PET/MR ScannerJournal of Nuclear Medicine, 2011
- Magnetic susceptibility mapping of brain tissue in vivo using MRI phase dataMagnetic Resonance in Medicine, 2009
- The potential of PET/MR for brain imagingEuropean Journal of Nuclear Medicine and Molecular Imaging, 2008
- Artifacts and pitfalls in diffusion MRIJournal of Magnetic Resonance Imaging, 2006
- Quantification and reduction of ghosting artifacts in interleaved echo‐planar imagingMagnetic Resonance in Medicine, 1997