Standardization of flow cytometry and cell sorting to enable a transcriptomic analysis in a multi-site sarcoidosis study
Open Access
- 9 March 2023
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 18 (3) , e0281210
Abstract
The contribution and regulation of various CD4+ T cell lineages that occur with remitting vs progressive courses in sarcoidosis are poorly understood. We developed a multiparameter flow cytometry panel to sort these CD4+ T cell lineages followed by measurement of their functional potential using RNA-sequencing analysis at six-month intervals across multiple study sites. To obtain good quality RNA for sequencing, we relied on chemokine receptor expression to identify and sort lineages. To minimize gene expression changes induced by perturbations of T cells and avoid protein denaturation caused by freeze/thaw cycles, we optimized our protocols using freshly isolated samples at each study site. To accomplish this study, we had to overcome significant standardization challenges across multiple sites. Here, we detail standardization considerations for cell processing, flow staining, data acquisition, sorting parameters, and RNA quality control analysis that were performed as part of the NIH-sponsored, multi-center study, BRonchoscopy at Initial sarcoidosis diagnosis Targeting longitudinal Endpoints (BRITE). After several rounds of iterative optimization, we identified the following aspects as critical for successful standardization: 1) alignment of PMT voltages across sites using CS&T/rainbow bead technology; 2) a single template created in the cytometer program that was used by all sites to gate cell populations during data acquisition and cell sorting; 3) use of standardized lyophilized flow cytometry staining cocktails to reduce technical error during processing; 4) development and implementation of a standardized Manual of Procedures. After standardization of cell sorting, we were able to determine the minimum number of sorted cells necessary for next generation sequencing through analysis of RNA quality and quantity from sorted T cell populations. Overall, we found that implementing a multi-parameter cell sorting with RNA-seq analysis clinical study across multiple study sites requires iteratively tested standardized procedures to ensure comparable and high-quality results.Keywords
Funding Information
- National Heart, Lung, and Blood Institute (5R01HL136681)
This publication has 38 references indexed in Scilit:
- Validation of cell‐based fluorescence assays: Practice guidelines from the ICSH and ICCS – part V – assay performance criteriaCytometry Part B: Clinical Cytometry, 2013
- Quality assurance for polychromatic flow cytometry using a suite of calibration beadsNature Protocols, 2012
- EuroFlow standardization of flow cytometer instrument settings and immunophenotyping protocolsLeukemia, 2012
- Recommendations for the validation of flow cytometric testing during drug development: II assaysJournal of Immunological Methods, 2011
- Recommendations for the validation of flow cytometric testing during drug development: I instrumentationJournal of Immunological Methods, 2011
- Optimizing a Multicolor Immunophenotyping AssayClinics In Laboratory Medicine, 2007
- CXCL9 and 11 in patients with pulmonary sarcoidosis: a role of alveolar macrophagesClinical and Experimental Immunology, 2007
- Data quality assessment of ungated flow cytometry data in high throughput experimentsCytometry Part A, 2007
- Elevated Serum Levels of CXCL9/Monokine Induced by Interferon-γ and CXCL10/Interferon-γ-Inducible Protein-10 in Ocular SarcoidosisInvestigative Ophthalmology & Visual Science, 2006
- Instrument‐Dependent fluorochrome sensitivity in flow cytometric analysesCytometry, 1995