Zoom lens design using liquid lens for laparoscope
Authors:
- 16 January 2013
- journal article
- Published by Optica Publishing Group in Optics Express
- Vol. 21 (2) , 1751-1761
Abstract
Traditional laparoscopic optical systems consisting of about 30 lenses have low optical magnification. To magnify tissue during surgical operations, one must change from one laparoscope to another or use a magnifying adapter between the laparoscope and the sensor. Our work focuses on how to change the sag of a liquid lens while zooming from 1 × zoom, to 2 × , and 4 × in an optical design for a laparoscope. The design includes several lenses and two liquid lenses with variable focal lengths. A pair of laparoscopes for 3-D stereoscopy is placed within a tube 11 mm in diameter. The predicted depth resolution of tissue is 0.5 mm without interpolation at 4 × zoom.Keywords
This publication has 9 references indexed in Scilit:
- Microelectrofluidic iris for variable aperture2012
- An Optofluidic Concept for a Tunable Micro-irisJournal of Microelectromechanical Systems, 2010
- Fluidic lens laparoscopic zoom camera for minimally invasive surgeryJournal of Biomedical Optics, 2010
- Comparing endoscopic systems on two simulated tasksErgonomics, 2005
- Components for integrated poly(dimethylsiloxane) microfluidic systemsElectrophoresis, 2002
- Fabrication of microfluidic systems in poly(dimethylsiloxane)Electrophoresis, 1999
- The effect of a second-generation 3D endoscope on the laparoscopic precision of novices and experienced surgeonsSurgical Endoscopy, 1999
- High Resolution Imaging of the Upper Respiratory Tract with Optical Coherence Tomography: A Feasibility StudyAmerican Journal of Respiratory and Critical Care Medicine, 1998
- Surface tension of lithium bromide solutions with heat-transfer additivesJournal of Chemical & Engineering Data, 1991