Single‐breath analysis using a novel simple sampler and capillary electrophoresis with contactless conductometric detection
- 6 November 2014
- journal article
- research article
- Published by Wiley in Electrophoresis
- Vol. 36 (4) , 526-533
Abstract
The analysis of ionic content of exhaled breath condensate (EBC) from one single breath by CE with C4D is demonstrated for the first time. A miniature sampler made from a 2‐mL syringe and an aluminum cooling cylinder for collection of EBC was developed. Various parameters of the sampler that influence its collection efficiency, repeatability, and effect of respiratory patterns were studied in detail. Efficient procedures for the cleanup of the miniature sampler were also developed and resulted in significant improvement of sampling repeatability. Analysis of EBC was performed by CE‐C4D in a 60 mM MES/l‐histidine BGE with 30 μM CTAB and 2 mM 18‐crown‐6 at pH 6 and excellent repeatability of migration times (RSD < 1.3% (n = 7)) and peak areas (RSD < 7% (n = 7)) of 12 inorganic anions, cations, and organic acids was obtained. It has been shown that the breathing pattern has a significant impact on the concentration of the analytes in the collected EBC. As the ventilatory pattern can be easily controlled during single exhalation, the developed collection system and method provides a highly reproducible and fast way of collecting EBC with applicability in point‐of‐care diagnostics.Keywords
Funding Information
- Czech Science Foundation (P206/13/21919S)
This publication has 40 references indexed in Scilit:
- Portable Capillary Electrophoresis Instrument with Automated Injector and Contactless Conductivity DetectionAnalytical Chemistry, 2013
- Determination of fluoroacetate and fluoride in blood serum by capillary zone electrophoresis using capacitively coupled contactless conductivity detectionElectrophoresis, 2011
- Quantification of plasma lactate concentrations using capillary electrophoresis with contactless conductivity detectionElectrophoresis, 2011
- Increased metal concentrations in exhaled breath condensate of industrial welders2010
- Feasibility of a new method to collect exhaled breath condensate in pre‐school childrenPediatric Allergy and Immunology, 2010
- Efficacy of two breath condensersJournal of Clinical Laboratory Analysis, 2010
- Exhaled breath condensate: lessons learned from veterinary medicineJournal of Breath Research, 2009
- Comparison of Different Contactless Conductivity Detectors for the Determination of Small Inorganic Ions by Capillary ElectrophoresisElectroanalysis, 2006
- Comparison of two methods for exhaled breath condensate collectionAllergy, 2006
- The Aerocrine exhaled nitric oxide monitoring system NIOX is cleared by the US Food and Drug Administration for monitoring therapy in asthmaJournal of Allergy and Clinical Immunology, 2004