Preliminary characterization and fundamental properties of aerosols generated by a flow focusing pneumatic nebulizer
- 1 January 2004
- journal article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 19 (10) , 1340-1346
Abstract
A new pneumatic concentric nebulizer (flow focusing pneumatic nebulizer, FFPN), based on the capillary flow focusing effect which produces highly controllable fine aerosols, is presented for the first time for analytical applications. The key aerosol diagnostic data and analytical figures of merit obtained using ICP-AES are presented for the FFPN and compared with a commercially available glass concentric nebulizer (conical nebulizer, CK). To perform a fair comparison, both nebulizers have been run under their own standard working conditions. Under all the conditions studied, the FFPN produces a finer primary aerosol than the conventional pneumatic nebulizer. At a solution uptake rate of 2.0 mL min−1 and a nebulizer gas flow rate of 1.0 L min−1 for the CK, and at a solution uptake rate of 0.2 mL min−1 and a nebulizer gas flow rate of 0.7 L min−1 for the FFPN, the median diameter (D50) values of the primary aerosol are 24.64 μm and 3.82 μm with the CK and FFPN, respectively. The tertiary aerosols of the CK and FFPN studied do not show a substantial difference, the D50 values, under the same operating conditions stated above, being 2.89 μm and 2.87 μm for the CK and FFPN, respectively. Under most of the studied conditions, all the primary aerosols produced by FFPN are contained on droplets smaller than 10 μm. For example, under the above experimental conditions, nearly 22% and 98% of primary aerosol volume is contained on droplets having sizes less than 10 μm for the CK and FFPN, respectively. However, approximately all the tertiary aerosol volume is contained on droplets smaller than this diameter with both nebulizers. This is the result of strong filtering action of the cyclonic spray chamber used, with a cut-off diameter of 15 μm, when CK is used. In general, sensitivities, precision and limits of detection with the FFPN are similar or better than those of the CK, yet with a sample uptake rate of only one-tenth the CK rate.Keywords
This publication has 37 references indexed in Scilit:
- Developments with the oscillating capillary nebulizer—effects of spray chamber design, droplet size and turbulence on analytical signals and analyte transport efficiency of selected biochemically important organoselenium compoundsJournal of Analytical Atomic Spectrometry, 2002
- Sonic Spray Nebulizer for Inductively Coupled Plasma Atomic Emission SpectrometryAnalytical Sciences, 1999
- Comparison of characteristics and limits of detection of pneumatic micronebulizers and a conventional nebulizer operating at low uptake rates in ICP-AESJournal of Analytical Atomic Spectrometry, 1999
- Acid effects in inductively coupled plasma atomic emission spectrometry with different nebulizers operated at very low sample consumption ratesJournal of Analytical Atomic Spectrometry, 1998
- Inductively Coupled Plasma Optical Emission Spectrometry Using Nebulizers with Widely Different Sample Consumption RatesAnalytical Chemistry, 1994
- Performance of the New Single-Bore High-Pressure Pneumatic Nebulizer (SBHPPN) in Plasma Atomic Emission SpectrometryApplied Spectroscopy, 1994
- High-efficiency nebulizer for argon inductively coupled plasma mass spectrometryJournal of Analytical Atomic Spectrometry, 1994
- Elemental speciation by anion exchange and size exclusion chromatography with detection by inductively coupled plasma mass spectrometry with direct injection nebulizationAnalytical Chemistry, 1993
- Evolution of drop size distributions for pneumatically generated aerosols in inductively coupled plasma-atomic emission spectrometrySpectrochimica Acta Part B: Atomic Spectroscopy, 1990
- High pressure nebulization: a new way of sample introduction for atomic spectroscopyAnalytical and Bioanalytical Chemistry, 1988