An optimum method for pulsed high intensity focused ultrasound treatment of large volumes using the InSightec ExAblate® 2000 system
Authors:
- 12 October 2010
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 55 (21) , 6395-6410
Abstract
Pulsed high intensity focused ultrasound (pHIFU) is a method for delivering ultrasound to tissue while avoiding high temperatures. The technique has been suggested for non-destructively enhancing local uptake of drugs. Side effects include thermal necrosis; therefore, real-time monitoring of tissue temperature is advantageous. This paper outlines a method for improving the treatment efficiency of pHIFU using the MR image-guided InSightec ExAblate® 2000 system, an ultrasound system integrated into a whole body human MRI scanner with the ability to measure temperature at the treatment location in near real time. Thermal measurements obtained during treatment of a tissue phantom were used to determine appropriate heating parameters, and compared to in vivo treatment of rabbit muscle. Optimization of the treatment procedure and ultrasound transducer steering patterns was then conducted with the goal of minimizing treatment time while avoiding overheating. The optimization was performed on the basis of approximate solutions to the standard bioheat equation. The commercial system software of the Exablate® system was modified to assist in this optimization. Depending on the size of the treatment volume, the presented results demonstrate that it is possible to use the technique described to cut treatment times significantly, up to one-third of that required by the current standard treatment cycle.Keywords
This publication has 15 references indexed in Scilit:
- Pulsed High Intensity Focused Ultrasound Mediated Nanoparticle Delivery: Mechanisms and Efficacy in Murine MuscleUltrasound in Medicine & Biology, 2009
- A Clinical Review of Focused Ultrasound Ablation With Magnetic Resonance GuidanceUltrasound Quarterly, 2008
- Therapeutic applications of ultrasoundProgress in Biophysics and Molecular Biology, 2007
- Delivery of Liposomal Doxorubicin (Doxil) in a Breast Cancer Tumor Model: Investigation of Potential Enhancement by Pulsed-High Intensity Focused Ultrasound ExposureAcademic Radiology, 2006
- Pulsed High-Intensity Focused Ultrasound Enhances Systemic Administration of Naked DNA in Squamous Cell Carcinoma Model: Initial ExperienceRadiology, 2005
- Thermal dosimetry of a focused ultrasound beam in vivo by magnetic resonance imagingMedical Physics, 1999
- Pennes’ 1948 paper revisitedJournal of Applied Physiology, 1998
- In vivo target-specific delivery of macromolecular agents with MR-guided focused ultrasound.Radiology, 1997
- A precise and fast temperature mapping using water proton chemical shiftMagnetic Resonance in Medicine, 1995
- Thermal dose determination in cancer therapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1984