Direct aperture optimization for a variable aperture collimator for intensity-modulated radiation therapy
Author:
- 13 February 2004
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 49 (5) , N47-N55
Abstract
In this note a technique is described for direct aperture optimization of components deliverable by a variable aperture collimator (VAC) for intensity-modulated radiation therapy. The first key result found was that, provided a large number of VAC components were selected for optimization, the resulting fluence profiles and the dose distribution were quite similar, but not identical, to the outcome of a direct inverse-planning algorithm in which the fluence of each bixel was individually adjusted during the iteration process. A second key feature is the ability to be able to construct highly modulated beams from a quite limited number of such components. It was shown that, when the number fell from 300 to 30, a recognizable conformal dose distribution was still obtainable although poorer. The conclusion was that the technique has the flexibility to cope with optimizing any specified number of VAC components and to observe the effect on the dose distribution of reducing this number.Keywords
This publication has 10 references indexed in Scilit:
- The physical basis of IMRT and inverse planningBritish Journal of Radiology, 2003
- Intensity-modulated radiation therapy using a variable-aperture collimatorPhysics in Medicine & Biology, 2003
- Direct aperture optimization: A turnkey solution for step‐and‐shoot IMRTMedical Physics, 2002
- Intensity-modulated radiotherapy by means of static tomotherapy: A planning and verification studyMedical Physics, 1997
- Realization and verification of three-dimensional conformal radiotherapy with modulated fieldsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1994
- Dynamic X-ray compensation for conformal radiotherapy by means of multi-leaf collimationRadiotherapy and Oncology, 1994
- Generation of arbitrary intensity profiles by dynamic jaws or multileaf collimatorsMedical Physics, 1994
- X-ray field compensation with multileaf collimatorsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1994
- An analytical solution for the dynamic control of multileaf collimatorsPhysics in Medicine & Biology, 1994
- Optimization of conformal radiotherapy dose distributions by simulated annealing: II. Inclusion of scatter in the 2D techniquePhysics in Medicine & Biology, 1991