Light hazards to the patient’s retina from ophthalmic instruments
Authors:
- 1 June 1991
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 30 (16) , 2187-2196
Abstract
The spectral radiance and irradiance of several common ophthalmic instruments were measured. The respective spectral retinal irradiance a patient might receive was calculated. To evaluate the potential light hazard to the patient’s retina, we calculated the temperature rise resulting from light absorption at the fundus, compared retinal irradiance with threshold values for photochemical retinal damage, and correlated them with safety guidelines. A thermal hazard to the patient’s retina can result from operation microscopes if light output is maximal and exposure prolonged. Comparison of light exposures from different ophthalmic instruments with thresholds for photochemical-induced lesions showed that one has to be more concerned about the actinic effect of light exposure. At maximum lamp voltage, the light exposure from operation microscopes reaches thresholds for photochemical damage even after a few minutes. Other ophthalmic instruments present no particular hazard if they are used at common exposure durations. Application of laser safety guidelines confirm these results. In some cases, light from operation microscopes exceeds the safety margin in <1 min.Keywords
This publication has 14 references indexed in Scilit:
- Observations on Monkey Eyes Exposed to Light from an Operating MicroscopeOphthalmology, 1983
- Action Spectrum for Retinal Injury from Near-Ultraviolet Radiation in the Aphakic MonkeyAmerican Journal of Ophthalmology, 1982
- Potential hazards from specific ophthalmic devicesVision Research, 1980
- Retinal Damage from LightAmerican Journal of Ophthalmology, 1979
- The damaging effects of light on the retina. Empirical findings, theoretical and practical implicationsSurvey of Ophthalmology, 1978
- Retinal damage secondary to chronic light exposureDocumenta Ophthalmologica, 1977
- Retinal sensitivity to damage from short wavelength lightNature, 1976
- Effects of intense visible radiation on the increment-threshold spectral sensitivity of the rhesus monkey eyeVision Research, 1975
- Photic Maculopathy Produced by the Indirect OphthalmoscopeAmerican Journal of Ophthalmology, 1972
- Ocular Spectral Characteristics as Related to Hazards from Lasers and Other Light SourcesAmerican Journal of Ophthalmology, 1968