Are new technologies always reducing patient doses in cardiac procedures?
Authors:
- A. TrianniATA. Trianni
- G. BernardiGBG. Bernardi
- R. PadovaniRPR. Padovani
- 1 December 2005
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 117 (1-3) , 97-101
Abstract
Dynamic digital flat-panel (FD) imaging technology is characterised by a higher sensitivity and image quality compared to image intensifier (II) technology. In this study, an angiography system based on II and one based on FD were compared with respect to system performance and impact of patient dose in interventional cardiology. When entrance surface air kerma rates are measured, the FD system requires a reduced dose rate, of up to 40%. For coronary angiography (CA), fluoroscopy time (FT) and dose–area product (DAP) were 4.3 ± 5.0 min and 31.2 ± 30.2 Gy cm2 on the II system and 4.4 ± 3.8 min and 33.4 ± 19.2 Gy cm2 with the FD system. For percutaneous transluminal coronary angiography, FT and DAP were 11.4 ± 10.7 min and 52.1 ± 45.0 Gy cm2 on II and 10.7 ± 8.7 min and 66.9 ± 54.4 Gy cm2 on DF. Data comparison suggests that reduced entrance dose rates do not automatically imply a reduction of patient dose in clinical practice.Keywords
This publication has 9 references indexed in Scilit:
- Dose performance evaluation of a charge coupled device and a flat-panel digital fluoroscopy system recently installed in an interventional cardiology laboratoryRadiation Protection Dosimetry, 2004
- Comparison of a conventional and a flat-panel digital system in interventional cardiology proceduresBritish Journal of Radiology, 2004
- Design and performance of a high-quality cardiac flat detector2002
- Progress report on the performance of real-time selenium flat-panel detectors for direct x-ray imaging2002
- Development and evaluation of a digital radiography system using a large-area flat-panel detector2002
- Fast imaging of a 41-cm amorphous silicon flat-panel detector for radiographic applications2001
- Experimental evaluation of a novel CdZnTe flat-panel x-ray detector for digital radiography and fluoroscopy2001
- Performance of a flat-panel cardiac detector2001
- Performance of a 41×41‐cm amorphous silicon flat panel x‐ray detector for radiographic imaging applicationsMedical Physics, 2000