Front-end ASIC for a GEM based time projection chamber
Authors:
- C. ThornCTC. Thorn
- Bo YuBYBo Yu
- 16 August 2004
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 51 (4) , 1312-1317
Abstract
A time projection chamber for experiments with the laser electron gamma source is being developed. It is composed of a can and a single-ended, dual-stage gas electron multiplier, with associated anode plane pixellated into about 8000 pads. The front-end electronics must provide energy, timing, and address information from those pads involved in measuring each track. For center of gravity determination this information must be sampled from the above-threshold pad and from the two adjacent ones. An efficient scheme for readout of the /spl ap/8000 front-end channels between each measurement cycle is also required. A 32-channel front-end application-specific integrated circuit has been developed to serve this detector. Fabricated in 0.25 /spl mu/m CMOS technology, it dissipates 41 mW. Each channel implements a low noise charge preamplifier with continuous reset of new concept, shaping amplifier with bandgap referenced baseline stabilizer, single threshold discriminator, dual-phase peak detector, timing detector, and logic for neighbor enabling. The readout process is based on token passing and flag.Keywords
This publication has 10 references indexed in Scilit:
- Development of a high-rate high-resolution detector for EXAFS experimentsIEEE Transactions on Nuclear Science, 2003
- Study of GEM characteristics for application in a MicroTPCIEEE Transactions on Nuclear Science, 2003
- Analog front-end cell designed in a commercial 0.25 /spl mu/m process for the ATLAS pixel detector at LHCIEEE Transactions on Nuclear Science, 2002
- Analog CMOS peak detect and hold circuits. Part 2. The two-phase offset-free and derandomizing configurationNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2002
- A CMOS fully compensated continuous reset systemIEEE Transactions on Nuclear Science, 2000
- A CMOS baseline holder (BLH) for readout ASICsIEEE Transactions on Nuclear Science, 2000
- A CMOS detector leakage current self-adaptable continuous reset system: Theoretical analysisNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1999
- GEM: A new concept for electron amplification in gas detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997
- Design of Low-Voltage, Low-Power Operational Amplifier Cells1996